[SENSe:]VOLTage[:DC]:RESolution
Syntax
[SENSe:]VOLTage[:DC]:RESolution {<resolution>|MIN|MAX|DEF} [, (@<ch_list>)]
[SENSe:]VOLTage[:DC]:RESolution? [{(@<ch_list>)|MIN|MAX}]
Description
This command selects the measurement resolution for dc voltage measurements on the specified channels. Specify the resolution in the same units as the selected measurement function, not in number of digits. If you omit the optional <ch_list> parameter, this command applies to the internal DMM, independent of any channels or a scan list.
Used With:
34921A through 34925A Multiplexer Modules
Parameters
Name |
Type |
Range of Values |
Default Value |
<resolution> |
Numeric |
Desired resolution in volts. |
0.000003 x Range (1 PLC) |
<ch_list> |
Numeric |
One or more channels in the form (@sccc). |
If <ch_list> is omitted, this command applies to the internal DMM. |
Remarks
For the <resolution> parameter, you can substitute MIN or MAX for a numeric value. MIN selects the smallest value accepted, which gives the highest resolution; MAX selects the largest value accepted, which gives the least resolution.
If you omit the optional <ch_list> parameter, the configuration applies to the internal DMM, independent of any channels or a scan list (any existing scan list is not redefined). An error is generated if the internal DMM is not installed or is disabled.
Setting the resolution also sets the integration time for the measurement. The following table shows the relationship between integration time, measurement resolution, number of digits, and number of bits.
Integration Time |
Resolution |
Digits |
Bits |
0.02 PLC |
< 0.0001 x Range |
4 |
15 |
0.2 PLC |
< 0.00001 x Range |
5 |
18 |
1 PLC |
< 0.000003 x Range |
5 |
20 |
2 PLC |
< 0.0000022 x Range |
6 |
21 |
10 PLC |
< 0.000001 x Range |
6 |
24 |
20 PLC |
< 0.0000008 x Range |
6 |
25 |
100 PLC |
< 0.0000003 x Range |
6 |
26 |
200 PLC |
< 0.00000022 x Range |
6 |
26 |
You can also set the integration time by specifying an aperture time (see [SENSe:]VOLTage[:DC]:APERture command). However, note that specifying integration time using NPLCs executes faster and offers better noise rejection characteristics for values of NPLC greater than 1.
The CONFigure, MEASure?, [SENSe:]VOLTage[:DC]:NPLC, and [SENSe:]VOLTage[:DC]:RESolution commands disable the aperture time mode (these commands select an integration time in number of power line cycles). You can use the [SENSe:]VOLTage[:DC]:APERture:ENABled? command to query whether the aperture mode is currently enabled or disabled.
The instrument sets the resolution to 1 PLC after a Factory Reset (*RST command). An Instrument Preset (SYSTem:PRESet command) or Card Reset (SYSTem:CPON command) does not change the setting.
Return Format
The query command returns the resolution in the form "+1.00000000E-03" for each channel specified. Multiple responses are separated by commas. Note that if the aperture mode is currently enabled (see [SENSe:]VOLTage[:DC]:APERture:ENABled? command), the value returned will represent the previous value specified, not the actual integration time.
Examples
The following command sets the measurement resolution to 1 mV on channels 3 and 13 in slot 1.
VOLT:DC:RES 1E-03,(@1003,1013)
The following query returns the resolution selected on channels 3 and 13 in slot 1.
VOLT:DC:RES? (@1003,1013)
VOLT:DC:APER:ENAB? !Verify
that aperture mode is disabled ("0")
Typical Response: |
+1.00000000E-03,+1.00000000E-03 |