[SENSe:]TEMPerature:APERture:ENABled?

34980A

[SENSe:]TEMPerature:APERture:ENABled?

Syntax

[SENSe:]TEMPerature:APERture:ENABled? [(@<ch_list>)]

Description

This command queries whether the aperture mode is enabled or disabled for temperature measurements on the specified channels. If you omit the optional <ch_list> parameter, this command applies to the internal DMM, independent of any channels or a scan list.

The following table shows which temperature transducers are supported by each of the multiplexer modules.

 

Thermocouple

RTD
2-Wire

RTD
4-Wire

Thermistor

 

34921A
Armature
Multiplexer

 Yes1

Yes

Yes

Yes

 

34922A
Armature Multiplexer

Yes2

Yes

Yes

Yes

 

34923A
Reed Multiplexer
(2-Wire)

Yes2,3

Yes4

Yes4

Yes5

 

34923A
Reed Multiplexer
(1-Wire)

Yes2,3

Yes4

No

Yes5

 

34924A
Reed Multiplexer

Yes2,3

Yes4

Yes4

Yes5

 

34925A
FET Multiplexer
(2-Wire)

Yes2

No

Yes5

No

 

34925A
FET Multiplexer
(1-Wire)

Yes2

No

No

No

 

1 Optional 34921T Terminal Block is required for thermocouple measurements with built-in internal reference junction.
2
A fixed or external reference junction temperature is required for thermocouple measurements with this module.
3
Impact of higher offset voltage specification (< 50 µV) must be taken into consideration.
4
1 kΩ or higher range used unless 100Ω series resistors are bypassed on module.
5
10 kΩ or higher range used for loads over approximately 300Ω due to series resistance of FET channels.

Parameters

Name

Type

Range of Values

Default Value

<ch_list>

Numeric

One or more channels in the form (@sccc).

If <ch_list> is omitted, this command applies to the internal DMM.

Remarks

  • The CONFigure, MEASure?, and [SENSe:]TEMPerature:NPLC commands disable the aperture mode (these commands select an integration time in number of power line cycles).

Return Format

The query command returns "0" is the aperture mode is disabled or "1" if the aperture mode is enabled for each channel specified. Multiple responses are separated by commas.

Example

The following command returns the state of the aperture mode on channels 3 and 13 in slot 1.

TEMP:APER:ENAB? (@1003,1013)

Typical Response:  1,1

See Also

CONFigure:TEMPerature

[SENSe:]TEMPerature:APERture

[SENSe:]TEMPerature:NPLC