OUTPut:ALARm<n>:SEQuence?

34980A

OUTPut:ALARm<n>:SEQuence

Syntax

OUTPut:ALARm<n>:SEQuence?

Description

This command returns the name of the sequence currently associated with the specified alarm.

Used With:

  • 34921A through 34925A Multiplexer Modules

  • 34950A Digital I/O Module (digital input and counter channels only)

  • 34952A Multifunction Module (digital input and totalizer channels only)

Parameters

Name

Type

Range of Values

Default Value

<n>

Discrete

{1|2|3|4}

This is a required parameter

Remarks

  • You can assign multiple channels to any of the four available alarms (numbered 1 through 4, see OUTPut:ALARm<n>:SOURce command). For example, you can configure the instrument to generate an alarm on the Alarm 1 output when a limit is exceeded on any of channels 1003, 2005, or 3010. You cannot, however, assign alarms on a specific channel to more than one alarm number.

  • A Factory Reset (*RST command) will remove all alarm associations. Redefining or deleting a sequence will also remove all alarm associations.

Return Format

This command returns the sequence name currently associated with the specified alarm. If no sequence has been associated, "NONE" is returned.   

Example

The following query returns the sequence name currently associated with Alarm 1.

OUTP:ALAR1:SEQ?

Typical Response:  MYSEQ_1

See Also

OUTPut:ALARm<n>:SOURce

ROUTe:SEQuence:TRIGger:SOURce