[SENSe:]RESistance:APERture:ENABled?

34980A

[SENSe:]RESistance:APERture:ENABled?

Syntax

[SENSe:]RESistance:APERture:ENABled? [(@<ch_list>)]

Description

This command queries whether the aperture mode is enabled or disabled for 2-wire resistance measurements on the specified channels. If you omit the optional <ch_list> parameter, this command applies to the internal DMM, independent of any channels or a scan list.

Used With:

  • 34921A through 34925A Multiplexer Modules

Parameters

Name

Type

Range of Values

Default Value

<ch_list>

Numeric

One or more channels in the form (@sccc).

If <ch_list> is omitted, this command applies to the internal DMM.

Remarks

  • Since 2-wire and 4-wire resistance are related functions, changing a measurement parameter for one function will also change the corresponding parameter for the other function (see [SENSe:]FRESistance:APERture:ENABled? command).

  • The CONFigure, MEASure?, [SENSe:]RESistance:NPLC, and [SENSe:]RESistance:RESolution commands disable the aperture mode (these commands select an integration time in number of power line cycles).

Return Format

The query command returns "0" is the aperture mode is disabled or "1" if the aperture mode is enabled for each channel specified. Multiple responses are separated by commas.

Example

The following command returns the state of the aperture mode on channels 3 and 13 in slot 1.

RES:APER:ENAB? (@1003,1013)

Typical Response:  1,1

See Also

CONFigure:RESistance

[SENSe:]RESistance:APERture

[SENSe:]RESistance:NPLC