INSTrument:DMM:DISConnect
Syntax
INSTrument:DMM:DISConnect
INSTrument:DMM:DISConnect?
Description
This command disconnects the internal DMM from the Analog Buses (ABus1 and ABus2).
Remarks
With the internal DMM disconnected, you can still configure DMM measurements and channel-based measurements; however, the internal DMM will not be reconnected until you send the INSTrument:DMM:CONNect command or a Factory Reset (*RST command).
The internal DMM setting is stored as part of the instrument state by the *SAV command.
The internal DMM is an optional assembly for the Agilent 34980A. This command will generate an error if the internal DMM is disabled (see INSTrument:DMM[:STATe] command) or not installed in the mainframe.
The connection setting is stored in volatile memory and will be connected (OFF) when power is turned off or after a Factory Reset (*RST command).
Return Format
The query command returns "0" (connected) or "1" (disconnected).
Examples
The following command disconnects the internal DMM from ABus1 and ABus2.
INST:DMM:DISC
The following query returns the status of the internal DMM connection.
INST:DMM:DISC?
Typical Response: 1