DIAGnostic:RELay:CYCLes:CLEar
Syntax
DIAGnostic:RELay:CYCLes:CLEar (@<ch_list>)
Description
This command resets the cycle count on the specified channels. In addition to channel relays, you can also clear the cycle count on bank relays and Analog Bus relays.
Parameters
Name |
Type |
Range of Values |
Default Value |
<ch_list> |
Numeric |
One or more channels in the form (@sccc). |
This is a required parameter |
Remarks
The instrument must be unsecured in order to reset the cycle count. See the CALibration:SECure:STATe command for more information on unsecuring the instrument.
The 34923A, 34924A, and 34933A modules can be configured for 2-wire (differential) or 1-wire (single ended) measurements (see SYSTem:MODule:WIRE:MODE command). Since two coils are required to drive each channel relay in the 2-wire mode, the count on both coils will be cleared. To clear the count on individual coils, reconfigure the module for the 1-wire mode (a power cycle is required) and clear the count.
On the RF Multiplexer modules (34941A, 34942A), each bank consists of two leaf relays and one tree relay (see DIAGnostic:RELay:CYCLes? command). Clearing the cycle count on a specific channel will clear the count on all three relays in the corresponding bank.
For paired operations on the 34945A Microwave Switch/Attenuator Driver (see ROUTe:CHANnel:DRIVe:PAIRed[:MODE] command), clearing the count on the lower channel of the pair will not clear the count on the upper channel.
Example
The following command clears the cycle count on channels 3 and 13 in slot 1.
DIAG:REL:CYCL:CLE (@1003,1013)