[SENSe:]TEMPerature:RJUNction[:INTernal]?

34980A

[SENSe:]TEMPerature:RJUNction[:INTernal]?

Syntax

[SENSe:]TEMPerature:RJUNction[:INTernal]? (@<ch_list>)

Description

This command reads the internal reference junction temperature for thermocouple measurements on the specified channels (useful only for an internal reference source).

Used With:

  • 34921A Armature Multiplexer (with optional 34921T Terminal Block)

Parameters

Name

Type

Range of Values

Default Value

<ch_list>

Numeric

One or more channels in the form (@sccc).

This is a required parameter

Remarks

  • This command is useful only for an internal reference source (see TEMPerature:TRANsducer:TCouple:RJUNction:TYPE command).

  • An error is generated if the optional 34921T Terminal Block is not attached to the module (the transducer is located on the terminal block).

  • The command always returns a temperature in degrees Celsius regardless of the temperature units currently selected (see UNIT:TEMPerature command).

Return Format

The query command returns the internal reference temperature in the form "+0.00000000E+00" for each channel specified (always in degrees Celsius). Multiple responses are separated by commas.

Example

The following query reads the internal reference junction temperature (always in degrees Celsius) on channels 3 and 13 in slot 1.

TEMP:RJUN:INT? (@1003,1013)

Typical Response:  +2.91170000E+01,+3.00230000E+01

See Also

CONFigure:TEMPerature

TEMPerature:TRANsducer:TCouple:RJUNction:TYPE

UNIT:TEMPerature