[SENSe:]TEMPerature:RJUNction[:INTernal]?
Syntax
[SENSe:]TEMPerature:RJUNction[:INTernal]? (@<ch_list>)
Description
This command reads the internal reference junction temperature for thermocouple measurements on the specified channels (useful only for an internal reference source).
Used With:
34921A Armature Multiplexer (with optional 34921T Terminal Block)
Parameters
Name |
Type |
Range of Values |
Default Value |
<ch_list> |
Numeric |
One or more channels in the form (@sccc). |
This is a required parameter |
Remarks
This command is useful only for an internal reference source (see TEMPerature:TRANsducer:TCouple:RJUNction:TYPE command).
An error is generated if the optional 34921T Terminal Block is not attached to the module (the transducer is located on the terminal block).
The command always returns a temperature in degrees Celsius regardless of the temperature units currently selected (see UNIT:TEMPerature command).
Return Format
The query command returns the internal reference temperature in the form "+0.00000000E+00" for each channel specified (always in degrees Celsius). Multiple responses are separated by commas.
Example
The following query reads the internal reference junction temperature (always in degrees Celsius) on channels 3 and 13 in slot 1.
TEMP:RJUN:INT? (@1003,1013)
Typical Response: +2.91170000E+01,+3.00230000E+01
See Also
CONFigure:TEMPerature