Agilent 34924A Module Summary
70-Channel Reed Multiplexer
Simplified Schematic
This module uses a Safety Interlock feature that prevents connections to the Analog Buses if no terminal block or cable is connected. |
Due to power dissipation issues, there is a limit to the number of relays that can be closed at a time on this module (you cannot close all channels simultaneously). See the ROUTe:CLOSe command for more information. |
Wiring Log
A wiring log is available to make it easy to document your wiring configuration for this module. You can open the file in Microsoft® Excel or Adobe® Acrobat® format.
34924A Wiring Log: Excel format Adobe PDF format
SCPI Commands Used
Key Specifications
See the Agilent 34921A-34925A Low Frequency Multiplexer Modules User's Guide for detailed operating information.
Max Voltage |
±150 V |
Max Current |
0.5 A |
Bandwidth |
25 MHz |
Scanning Speed |
500 chan/sec |
Thermal Offset |
< 50 µV |
Note: Product specifications are subject to change without notice.