SYSTem:SECurity:IMMediate
Syntax
SYSTem:SECurity:IMMediate
Description
This command clears all instrument memory (except calibration parameters, I/O hardware addresses, and boot parameters) and then cycles power to the instrument. This command is typically used to clear all memory before removing the instrument from a secured area.
This command is not recommended for use in routine applications because of the possibility of unintended loss of data. |
Remarks
This command initializes all instrument and module settings to their Factory Reset (*RST command) values.
This command deletes all user-defined state information, channel labels, and I/O settings such as the GPIB and LAN address information.
The only non-volatile settings that are not deleted are the calibration parameters (calibration string, secure code, etc.), I/O hardware address information (Ethernet MAC address and instrument serial number), and the instrument's boot parameters.
Example
The following command clears all instrument memory.
SYST:SEC:IMM