[SENSe:]PERiod:VOLTage:RANGe

34980A

[SENSe:]PERiod:VOLTage:RANGe

Syntax

[SENSe:]PERiod:VOLTage:RANGe {<voltage_range>|MIN|MAX|DEF} [, (@<ch_list>)]

[SENSe:]PERiod:VOLTage:RANGe? [{(@<ch_list>)|MIN|MAX}]

Description

This command selects the voltage range for period measurements on the specified channels. If you omit the optional <ch_list> parameter, this command applies to the internal DMM, independent of any channels or a scan list.

Used With:

  • 34921A through 34925A Multiplexer Modules

Parameters

Name

Type

Range of Values

Default Value

<voltage_range>

Discrete

Desired range in volts:

100 mV (MIN)
1 V
10 V
100 V
300 V (MAX)

This is a required parameter

<ch_list>

Numeric

One or more channels in the form (@sccc).

If <ch_list> is omitted, this command applies to the internal DMM.

Remarks

  • If you omit the optional <ch_list> parameter, the configuration applies to the internal DMM, independent of any channels or a scan list (any existing scan list is not redefined). An error is generated if the internal DMM is not installed or is disabled.

  • Selecting a discrete range will disable autoranging on the specified channels.

  • If the input signal is greater than can be measured on the selected range (manual ranging), the instrument gives an overload indication: "±OVLD" from the front panel or "±9.9E+37" from the remote interface.

  • Since frequency and period are related functions, changing a measurement parameter for one function will also change the corresponding parameter for the other function (see [SENSe:]FREQuency:VOLTage:RANGe command).

  • The instrument enables autoranging after a Factory Reset (*RST command). An Instrument Preset (SYSTem:PRESet command) or Card Reset (SYSTem:CPON command) does not change the setting.

Return Format

The query command returns the voltage range in the form "+1.00000000E+01" for each channel specified. Multiple responses are separated by commas.

Examples

The following command selects the 10 volt range for period measurements on channels 3 and 13 in slot 1.

PER:VOLT:RANG 10,(@1003,1013)

The following query returns the voltage range selected on channels 3 and 13 in slot 1.

PER:VOLT:RANG? (@1003,1013)

Typical Response:  +1.00000000E+01,+1.00000000E+01

See Also

CONFigure:PERiod

[SENSe:]PERiod:VOLTage:RANGe:AUTO