Agilent 34952A Module Summary

34980A

Agilent 34952A Module Summary

Multifunction Module with Digital I/O, Totalizer, and DAC

 

 

Simplified Block Diagram
 

 

 

Wiring Log

A wiring log is available to make it easy to document your wiring configuration for this module. You can open the file in Microsoft® Excel or Adobe® Acrobat® format.

34952A Wiring Log:     Excel format     Adobe PDF format

 

SCPI Commands Used
 

Digital I/O:

CONFigure:DIGital:DIRection

CONFigure:DIGital:WIDTh

MEASure:DIGital?

[SENSe:]DIGital:DATA[:<width>]?

[SENSe:]DIGital:DATA:BIT?

SOURce:DIGital:DATA[:<width>]

SOURce:DIGital:DATA:BIT

SOURce:DIGital:STATe

Totalizer:

CONFigure:TOTalize

MEASure:TOTalize?

[SENSe:]TOTalize:CLEar:IMMediate

[SENSe:]TOTalize:SLOPe

[SENSe:]TOTalize:THReshold[:MODE]

[SENSe:]TOTalize:TYPE

[SENSe:]TOTalize:DATA?

STATus:MODule:SLOT[n]:CONDition?

STATus:MODule:SLOT[n]:EVENt?

DAC:

SOURce:VOLTage[:LEVel]

 

Key Specifications

See the Agilent 34952A Multifunction Module User's Guide for detailed operating information.

Digital Input/Output

 

Vin(L)

< 0.8 V (TTL)

Vin(H)

> 2.0 V (TTL)

Vout(L)

< 0.8 V @ Iout = -400 mA

Vout(H)

> 2.4 V @ Iout = 1 mA

Vout(H) max

< 42 V with external pullup

Read/Write Speed

95/sec

 

Totalizer

 

Maximum Count

232 - 1  (4,294,967,295)

Totalize Input

100 kHz, rising or falling edge

Signal Level

1 Vpp (min), 42 Vpk (max)

Read Speed

85 readings/sec

 

DAC Output

 

DAC 1, 2

±12 V, non-isolated

Resolution

1 mV

Iout

10 mA max

Settling Time

1 ms to 0.01% of output

Note: Product specifications are subject to change without notice.