SYSTem:RMODule:RESet
Syntax
SYSTem:RMODule:RESet <slot>
Description
This command resets all 34945EXT remote modules connected to the 34945A Microwave Switch/Attenuator Driver in the specified slot. This command provides the only way to reset all 34945EXT remote modules in a given slot to a factory default state. The following table shows the reset state.
Attribute |
Reset State | |
At the 34945EXT Level: |
| |
|
OFF | |
|
OFF | |
|
1 | |
At the Bank Level: |
| |
|
OCOLlector | |
|
ON | |
|
5 mA | |
All Channels: |
| |
|
OFF | |
|
ON | |
|
15 ms | |
|
0.0 seconds | |
|
0.0 seconds | |
|
OPEN selected | |
|
OFF | |
|
NORMal |
Used With:
34945A Microwave Switch/Attenuator Driver
Parameters
Name |
Type |
Range of Values |
Default Value |
<slot> |
Discrete |
{1|2|3|4|5|6|7|8} |
This is a required parameter |
Remarks
Since this command disables both the channel drive source and channel drive boot state, the channels will remain inoperable until you re-enable the channel drive source using the ROUTe:RMODule:DRIVe:SOURce[:IMMediate] command (select the INTernal or EXTernal source).
With the SYSTem:RMODule:RESet command, the bank-level attributes are reset without regard to which distribution boards are installed. To reset the bank-level attributes on a specific 34945EXT remote module, see the ROUTe:RMODule:BANK:PRESet command; with this command, the bank-level attributes vary according to which distribution boards are installed.
The *RST and SYSTem:PRESet commands reset all channels on a remote module to their default states as specified by the ROUTe:CHANnel:DRIVe:OPEN:DEFault command and set all channel drives to the specified boot state (see ROUTe:RMODule:DRIVe:SOURce:BOOT command). In addition, when the *RST and SYSTem:PRESet commands are executed, the 34945A automatically re-evaluates all external remote module connections to determine what is connected, what is powered, etc. No other settings are changed.
The SYSTem:CPON command has nearly the same effect as the *RST and SYSTem:PRESet commands (see above). However, when the SYSTem:CPON command is executed, the 34945A does not re-evaluate the external remote module connections.
The SYSTem:RMODule:RESet command does not clear the relay cycle count on the specified slot (see DIAGnostic:RELay:CYCLes? command).
Example
The following command resets all 34945EXT remote modules connected to slot 3 to their factory default state. The bank-level attributes are reset without regard to which distribution boards are installed.
SYST:RMOD:RESET 3