ROUTe:SEQuence:ABORt
Syntax
ROUTe:SEQuence:ABORt
Description
This command aborts the sequence currently executing. If no sequence is current executing, the command is ignored (no error is generated).
Remarks
When the sequence is terminated, the resultant instrument state will be determined by how much of the sequence had been executed when the ABORt/Device Clear was received. An ABORt command (system abort) executed from within a sequence will not terminate the sequence. The *RST and SYSTem:PRESet commands will also abort a sequence execution prior to performing their own actions.
Example
The following command aborts the sequence currently executing.
ROUT:SEQ:ABOR