CALCulate:AVERage:CLEar
Syntax
CALCulate:AVERage:CLEar [(@<ch_list>)]
Description
This command clears all values from the statistics registers for each the specified channels. Each channel must be a channel that has been configured to be part of the scan list. If you omit the optional <ch_list> parameter, this command applies to the internal DMM, independent of any channels or a scan list.
Used With:
34921A through 34925A Multiplexer Modules
34950A Digital I/O Module (digital input and counter channels only)
34952A Multifunction Module (digital input and totalizer channels only)
Parameters
Name |
Type |
Range of Values |
Default Value |
<ch_list> |
Numeric |
One or more channels in the form (@sccc). |
If <ch_list> is omitted, this command applies to the internal DMM. |
Remarks
This command clears the minimum, maximum, average, count, and peak-to-peak values (but no readings are cleared from memory).
An error is generated if the internal DMM is disabled (see INSTrument:DMM[:STATe] command) or not installed in the mainframe.
The instrument clears the stored statistical data on all channels when a new scan is started, after a Factory Reset (*RST command), or after an Instrument Preset (SYSTem:PRESet command).
Return Format
This command has no query form.
Example
The following command clears the stored statistical data on channels 1 and 3 in slot 1.
CALC:AVER:CLE (@1001,1003)