*TRG

34980A

*TRG

Syntax

*TRG

Description

This command is used in conjunction with the TRIGger:SOURce BUS command to trigger the instrument from the remote interface.

Remarks

  • Use the TRIGger:SOURce command to select the BUS (software) trigger source.

  • After setting the trigger source, you must place the internal DMM in the "wait-for-trigger" state using the INITiate command. The *TRG command will not be accepted unless the internal DMM is in the "wait-for-trigger" state.

  • The *TRG command is not valid with the READ? command.

Example

The following command triggers the instrument.

TRIG:SOUR BUS
INIT
*TRG

See Also

INITiate

TRIGger:SOURce