*TRG
Syntax
*TRG
Description
This command is used in conjunction with the TRIGger:SOURce BUS command to trigger the instrument from the remote interface.
Remarks
Use the TRIGger:SOURce command to select the BUS (software) trigger source.
After setting the trigger source, you must place the internal DMM in the "wait-for-trigger" state using the INITiate command. The *TRG command will not be accepted unless the internal DMM is in the "wait-for-trigger" state.
The *TRG command is not valid with the READ? command.
Example
The following command triggers the instrument.
TRIG:SOUR BUS
INIT
*TRG