[SENSe:]TEMPerature:TRANsducer:TCouple:RJUNction:EXTernal?
Syntax
[SENSe:]TEMPerature:TRANsducer:TCouple:RJUNction:EXTernal?
Description
This command returns the most-recently stored external reference temperature value for thermocouple measurements. This temperature is used for all subsequent thermocouple measurements that specify an external reference source (see [SENSe:]TEMPerature:TRANsducer:TCouple:RJUNction:TYPE command).
Used With:
34921A Armature Multiplexer (with optional 34921T Terminal Block)
34922A Armature Multiplexer
34923A Reed Multiplexer (differential mode only)
34924A Reed Multiplexer
Remarks
To store a reference temperature, first configure a multiplexer channel for an RTD or thermistor measurement (see CONFigure:TEMPerature command). Then assign the measurement from that channel as the external reference using one of the following commands:
[SENSe:]TEMPerature:TRANsducer:FRTD:REFerence
[SENSe:]TEMPerature:TRANsducer:RTD:REFerence
[SENSe:]TEMPerature:TRANsducer:THERmistor:REFerence
When you initiate a measurement on the selected channel (see INITiate or READ? command), the acquired temperature is stored in volatile memory in the reference register. Subsequent thermocouple measurements use the stored temperature as their reference. The temperature remains in memory until you store a different value in the reference register or remove the mainframe power.At power-on, the reference register is loaded with "9.9E+37". Note that until you store a reference temperature, all thermocouple measurements that specify an external reference source will be returned as "9.9E+37".
Return Format
The query command returns the reference temperature in the form "+0.00000000E+00" (always in degrees Celsius).
Example
The following query reads reference register and returns the most-recently stored temperature measurement.
TEMP:TRAN:TC:RJUN:EXT?
Typical Response: +2.00000000E+01