INSTrument:DMM:CONNect

34980A

INSTrument:DMM:CONNect

Syntax

INSTrument:DMM:CONNect

INSTrument:DMM:CONNect?

Description

This command connects the internal DMM to the Analog Buses. When connected, the internal DMM is always connected to Analog Bus 1 (MEAS). For 4-wire measurements, the internal DMM is also connected to Analog Bus 2 (SENS).

Remarks

  • The internal DMM setting is stored as part of the instrument state by the *SAV command.

  • The internal DMM is an optional assembly for the Agilent 34980A. This command will generate an error if the internal DMM is disabled (see INSTrument:DMM[:STATe] command) or not installed in the mainframe.

  • The connection setting is stored in volatile memory and will be disconnected (OFF) when power is turned off or after a Factory Reset (*RST command).

Return Format

The query command returns "0" (disconnected) or "1" (connected).

Examples

The following command connects the internal DMM to ABus 1 and ABus2.

INST:DMM:CONN

The following query returns the status of the internal DMM connection.  

INST:DMM:CONN?

Typical Response:  1

See Also

INSTrument:DMM:DISConnect