Agilent 34921A Module Summary

34980A

Agilent 34921A Module Summary

40-Channel Armature Multiplexer with Low Thermal Offset

 

 

Simplified Schematic

 

This module uses a Safety Interlock feature that prevents connections to the Analog Buses if no terminal block or cable is connected.

 

 

Wiring Log

A wiring log is available to make it easy to document your wiring configuration for this module. You can open the file in Microsoft® Excel or Adobe® Acrobat® format.

34921A Wiring Log:     Excel format     Adobe PDF format

 

SCPI Commands Used

CONFigure Subsystem

MEASure? Subsystem

ROUTe:CLOSe

ROUTe:CLOSe:EXCLusive

ROUTe:OPEN

ROUTe:SCAN

SENSe Subsystem

 

Key Specifications

See the Agilent 34921A-34925A Low Frequency Multiplexer Modules User's Guide for detailed operating information.

Max Voltage

±300 V

Max Current

1 A

Bandwidth

45 MHz

Scanning Speed

100 chan/sec

Thermal Offset

< 3 µV

Note: Product specifications are subject to change without notice.