Automatic Trigger Delays
If you do not specify a trigger delay, the instrument selects a delay for you. The delay is determined by the function, range, integration time, and ac filter setting as shown below.
DC Voltage, Thermocouple, DC Current (for all ranges):
Integration Time |
Trigger Delay |
PLC > 1 |
2.0 ms |
PLC ≤ 1 |
1.0 ms |
Resistance, RTD, Thermistor (2- and 4-wire):
Range |
Trigger Delay (for PLC > 1) |
100Ω |
2.0 ms |
1 kΩ |
2.0 ms |
10 kΩ |
2.0 ms |
100 kΩ |
25 ms |
1 MΩ |
30 ms |
10 MΩ |
200 ms |
100 MΩ |
200 ms |
Range |
Trigger Delay (for PLC ≤ 1) |
100Ω |
1.0 ms |
1 kΩ |
1.0 ms |
10 kΩ |
1.0 ms |
100 kΩ |
20 ms |
1 MΩ |
25 ms |
10 MΩ |
200 ms |
100 MΩ |
200 ms |
AC Voltage, AC Current (for all ranges):
AC Filter |
Trigger Delay |
Slow (3 Hz) |
7.0 seconds |
Medium (20 Hz) |
1.0 second |
Fast (200 Hz) |
120 ms |
Frequency, Period:
AC Filter |
Trigger Delay |
Slow (3 Hz) |
600 ms |
Medium (20 Hz) |
300 ms |
Fast (200 Hz) |
100 ms |
Digital Input, Totalize:
Trigger Delay |
0 seconds |