Factory Reset State
The following tables show the state of the instrument after a *RST or SYSTem:CPON command is issued.
|
Measurement Configuration |
Factory Reset State |
|
Function |
DC Volts |
|
Range |
Autorange |
|
Resolution |
5 |
|
Integration Time |
1 PLC |
|
Input Resistance |
10 MΩ (fixed for all DCV ranges) |
|
Channel Labels |
No Change |
|
Channel Delay |
Automatic Delay |
|
Reading Format |
Reading Only (no units, channel, time) |
|
Sample Count |
1 Sample per Trigger |
|
Trigger Count |
1 Trigger |
|
Trigger Delay |
Automatic Delay |
|
Trigger Source |
Immediate |
|
Scanning Operations |
Factory Reset State |
|
Scan List |
Empty |
|
Reading Memory |
All Readings are Cleared |
|
Min, Max, and Average |
All Statistical Data is Cleared |
|
Sweep Count |
1 Sweep |
|
Trigger Interval |
1 Second |
|
Monitor in Progress |
Stopped |
|
Mx+B Scaling |
Factory Reset State |
|
Scaling State |
Off |
|
Gain Factor ("M") |
1 |
|
Offset Factor ("B") |
0 |
|
Scale Label |
Null String ( " " ) |
|
Alarm Limits |
Factory Reset State |
|
Alarm Queue |
Not Cleared |
|
Alarm State |
Off |
|
HI and LO Alarm Limits |
0 |
|
Alarm Output |
Alarm 1 |
|
Alarm Output Configuration |
Latched Mode |
|
Alarm Output State |
Output Lines are Cleared |
|
Alarm Output Slope |
Fail = Low |
|
Module Hardware |
Factory Reset State |
|
Multiplexer Modules: 34921A, 34922A, 34923A, 34924A, 34925A |
All Channels Open |
|
Matrix Modules: 34931A, 34932A, 34933A |
All Channels Open |
|
GP Modules: 34937A, 34938A, 34939A |
All Channels Open |
|
RF Modules: 34941A, 34942A |
Channels b01 and b03 Selected (b = Bank) |
|
Microwave Modules: 34945A, 34946A, 34947A |
34945A: All Channel Drives = Default |
|
System Control Modules: 34950A, 34951A, 34952A, 34959A |
34950A: DIO Ports = Input, Count = 0, |
|
System-Related Operations |
Factory Reset State |
|
Display State |
On |
|
Error Queue |
Errors Not Cleared |
|
Stored States |
No Change |
|
System Date |
No Change |
|
System Time |
No Change |
|
Temperature Units |
°C |