NI 654x Acquisition Termination
High-Impedance Load Acquisition Configuration
For acquisition operations, the NI 654x can only be used in a high-impedance load configuration because the input impedance of the NI 654x is set to 10 kΩ. The high-impedance load configuration is easy to drive since it does not present a significant DC load to the source and preserves the signal amplitude. It is very important that you follow the recommendations in this section to achieve the highest level of AC signal quality.
The same transmission line considerations discussed for the generation case are applicable to acquired signals. The input impedance of 10 kΩ implies a reflection coefficient Γt of 0.99, or nearly full reflection.
With all high-impedance load transmission lines, it is essential that you take care to match the source impedance of the transmission line to the characteristic impedance of the transmission line. The source matching in this configuration is particularly important, since there are significant reflections from the high-impedance load (input of NI 654x in this case).
You should take care to ensure that your transmission line has a characteristic impedance of as close to 50 Ω as possible.
Therefore, Zs (external source output impedance) should match Z0 = 50 Ω (the cable impedance) for 10 kΩ input configuration. To achieve this Zs = 50 Ω, you must determine the output impedance of your digital driver and add a series resistor as close as possible to the driver pin such that the output impedance of your buffer plus the value of the series resistor equal 50 Ω.
If you require 50 Ω termination for your acquisition application, consider using one of the NI 655x products.