Sweep

M9370A / M9371A / M9372A / M9374A / M9375A

Sweep Settings


A sweep is a series of consecutive data point measurements taken over a specified sequence of stimulus values. You can make the following sweep settings:

See Triggering and other 'Setup Measurements' topics

How to set Sweep Type

Using front-panel hardkey [softkey] buttons

Using Menus

  1. Press Sweep

  2. then [Sweep Type]

  1. Click Stimulus

  2. then Sweep

  3. then Sweep Type

 

Sweep Type dialog box help

Note: Sweep Settings are not applied until either OK or Apply is pressed.

Channel  The active channel when Sweep Type was selected. Sweep settings will be applied to this channel.

Sweep Type

Linear Frequency  Sets a linear frequency sweep that is displayed on a standard grid with ten equal horizontal divisions.

  • Start  Sets the beginning value of the frequency sweep.

  • Stop  Sets the end value of the frequency sweep.

  • Points  Sets the number of data points that the analyzer measures during a sweep. Range: 2 to 20001.(Default is 201).

  • Power - Sets the power level or the source. Learn more.

  • IF Bandwidth - Learn more.

Log Frequency  The source is stepped in logarithmic increments and the data is displayed on a logarithmic x-axis. This is usually slower than a continuous sweep with the same number of points.

  • Start  Sets the beginning value of the frequency sweep.

  • Stop  Sets the end value of the frequency sweep.

  • Points  Sets the number of data points that the analyzer measures during a sweep. Range: 2 to 20001. (Default is 201).

Power Sweep  Activates a power sweep at a single frequency that you specify.  Learn about power sweep

  • Start  Sets the beginning value of the power sweep.

  • Stop  Sets the end value of the power sweep.

  • CW Frequency  Sets the single frequency where the analyzer remains during the measurement sweep.

CW Time  Sets the analyzer to a single frequency, and the data is displayed versus time. Learn more.

  • CW Frequency   Sets the frequency where the analyzer remains during the measurement.

  • Sweep Time  Sets the duration of the measurement, which is displayed on the X-axis.

  • Points  Sets the number of data points that the analyzer measures during a sweep. Range: 2 to 20001.(Default is 201).

Segment Sweep  Sets the analyzer to sweep through user-defined sweep segments.  Learn how to make these settings.

  • Independent Power Levels  Check to set the source power level for each segment. Test port uncoupling is also allowed.

  • Independent IF Bandwidth  Check to set the IF bandwidth for each segment.

  • Independent Sweep Time  Check to set the duration of the measurement for each segment.

  • X-Axis Point Spacing  Check to scale the X-Axis to include only the segments. Learn more.

  • Allow Arbitrary Segments  Check to allow arbitrary frequencies (overlapped or reverse sweeps). Learn more

  • Show Table  Shows the table that allows you to create and edit segments.

  • Hide Table  Hides the segment table from the screen.

OK  Applies setting changes and closes the dialog box.

Apply  Applies setting changes and leaves the dialog box open to make more setting changes.

Cancel  Closes the dialog. Setting changes that have been made since the last Apply button click are NOT applied.
 

Power Sweep

A power sweep either increases or decreases source power in discrete steps. Power sweep is used to characterize power-sensitive circuits, with measurements such as gain compression.

In the Sweep Type dialog, specify Start power, Stop power, and CW Frequency. Power can be swept over any attainable range within the analyzer ALC range.

The remaining power settings apply in power sweep mode:

  • Test Port Power setting is not available.

  • Port Power can be coupled or uncoupled.

  • Attenuator Control is always Manual.

  • Power Slope (dB/GHz) is ignored (output frequency is CW).

  • Click Stimulus  then Sweep, then Number of Points to change the step size of the power sweep.

Notes:

  • Using a preference setting, you can specify whether to maintain source power at either the start power or stop power level at the end of a power sweep.

  • Power Sweep is optimized for speed. For highest measurement accuracy during a power sweep, it may be necessary to increase the Dwell Time to allow the source more time to settle.

Segment Sweep

Segment Sweep activates a sweep which consists of frequency sub-sweeps, called segments. For each segment you can define independent power levels, IF bandwidth, and sweep time.

Once a measurement calibration is performed on the entire sweep or across all segments, you can make calibrated measurements for one or more segments.

In segment sweep type, the analyzer does the following:

  • Sorts all the defined segments in order of increasing frequency

  • Measures each point

  • Displays a single trace that is a composite of all data taken

Restrictions for segment sweep:

  • The frequency range of a segment is not allowed to overlap the frequency range of any other segment.

  • The number of segments is limited only by the combined number of data points for all segments in a sweep.

  • The combined number of data points for all segments in a sweep cannot exceed the max number of data points per trace.

How to make segment sweep settings

Using front-panel
hardkey
[softkey] buttons

Using Menus

  1. Press Sweep

  2. then [Sweep Type]

  1. Click Stimulus

  2. then Sweep

  3. then Sweep Type

  4. then Segment

The Segment Table will be shown automatically when the Segment Table softkey menu is displayed. The Table will be hidden automatically after exiting the Segment Table menu. 

To display the Segment Table while using a menu other than the Segment Table menu, use the 4th softkey on the menu to control when the table is shown. The softkey offers three behaviors based on the softkey’s changing label.

  • Show Segment Table:  causes the Segment Table to appear but it will automatically hide when exiting the menu.

  • Hide Segment Table:  hides the Segment Table to appear but it will automatically reappear if the menu is exited then re-entered.

  • Pin Segment Table:  causes the Segment Table to appear and it will remain displayed after exiting the Segment Table menu.

To make the following menu settings available, you must first show the segment table.

From the menus, click Stimulus, then Sweep, then Segment Table, then Show Table.  

Then choose from the following:

Add Segment - adds a sweep segment at the end of the segments.

Insert Segment - adds a sweep segment before the selected segment.

Tip: You can also click the "down" arrow on your keyboard to quickly add many segments.

Delete Segment - removes the selected segment.

Delete All Segments - removes all segments.

Note: At least ONE segment must be ON or Sweep Type is automatically set to Linear.

To Modify an Existing Segment

The above graphic shows the Segment table with all independent settings selected, including source power uncoupled (two power settings).

STATE Click the box on the segment to be modified. Then use the up / down arrow to turn the segment ON or OFF.

START Sets start frequency for the segment. Click the box and type a value and the first letter of a suffix (KHz, Mhz, GHz). Or double-click the box  to select a value.

STOP  Sets stop frequency for the segment. Click the box and type a value and the first letter of a suffix (KHz, Mhz, GHz). Or double-click the box to select a value.

Note: The segment table truncates the frequency resolution. To verify the frequency resolution that you input, create a marker at the start or stop frequency settings.

POINTS  Sets number of data points for this segment. Type a value or double-click the box to select a value.

To set IFBW, Power, and Sweep Time independently for each segment:

  1. On the Sweep menu, click Sweep Type, then Segment Sweep.

  2. Check the appropriate Sweep Properties boxes

  3. Then click the box and type a value or double-click the box and select a value.

Note: If the following are NOT set, the entire sweep uses the channel IFBW, Power, and Time settings.

IFBW  Sets the IF Bandwidth for the segment.

POWER  Sets the Power level for the segment. You can also UNCOUPLE the test port power. See Power Coupling.

TIME  Sets the Sweep time for the segment.

X-Axis Point Spacing  - Segment Sweep ONLY

This feature affects how a segment trace is drawn on the screen.

How to select X-Axis Point Spacing

On the Sweep Type dialog box, click Segment Sweep

Then check X-Axis Point Spacing

  • Without X-axis point spacing, a multi-segment sweep trace can sometimes result in squeezing many measurement points into a narrow portion of the x-axis.

  • With X-axis point spacing, the x-axis position of each point is chosen so that all measurement points are evenly spaced along the x-axis.

For example, given the following two segments:

Without X-Axis Point Spacing

 

With X-Axis Point Spacing

Arbitrary Segment Sweep

This feature allows arbitrary frequencies to be entered into the segment sweep table. With this capability, segments can have:

  • overlapping frequencies.

  • the stop frequency less than the start frequency (reverse sweep).

How to enable Arbitrary Segment Sweep

  1. On the Sweep Type dialog box, click Segment Sweep

  2. Check Allow Arbitrary Segment Sweep

Notes:

    • Unusual results may occur when using arbitrary sweep segments with markers, display settings, limit lines, formatting, and some calibration features.

    • When Allow Arbitrary Segment is checked, X-axis point spacing is automatically turned ON.

Sweep Time

The analyzer automatically maintains the fastest sweep time possible with the selected measurement settings. However, you can increase the sweep time to perform a slower sweep.

How to set Sweep Time

Using front-panel
hardkey
[softkey] buttons

Using Menus

  1. Press Sweep

  2. then [Time]

  1. Click Stimulus

  2. then Sweep

  3. then Sweep Time

 

Time dialog box help

Sweep Time  Specifies the time the analyzer takes to acquire data for a sweep. The maximum sweep time of the analyzer is 86400 seconds or 1 day. Learn about other settings that affect sweep speed.

X-Axis Time  Set Start and Stop time to be displayed on the X-axis. These settings are NEVER available.

Note: Ignore CW time as labeled on the X-axis and marker readout. It is NOT at all accurate.

Sweep Setup

How to make Sweep Setup settings

Using front-panel
hardkey
[softkey] buttons

Using Menus

  1. Press Stimulus

  2. then [Sweep]

  3. then [Sweep Setup]

  1. Click Stimulus

  2. then Sweep

  3. then Sweep Setup

 

Sweep Setup dialog box help

Channel  Specifies the channel that the settings apply to.

Stepped Sweep  When checked (Stepped Sweep) the analyzer source is tuned, then waits the specified Dwell time, then takes response data, then tunes the source to the next frequency point. This is slower than Analog Sweep, but is more accurate when testing electrically-long devices.

When cleared (Analog Sweep) the analyzer takes response data AS the source is sweeping. The sweep time is faster than Stepped, but could cause measurement errors when testing electrically-long devices.

When the dialog check box is cleared, the analyzer could be in either Analog or Step mode. The mode can change from sweep to sweep. There is NO way to determine whether the analyzer is in Analog or Stepped Sweep. If you want to be sure what the current sweep mode is, then switch it to Stepped.

Stepped sweep is automatically selected for a number of reasons. Here are some of the reasons:

  • IF Bandwidth is at, or below, 1 kHz.

  • When step mode is a faster way to take the data.

Dwell Time  Specifies the time the source stays at each measurement point before the analyzer takes the data. Only applies to stepped sweep. The maximum dwell time is 100 seconds. See also Electrically Long Devices.

Sweep Delay  Specifies the time to wait just before acquisition begins for each sweep. This delay is in addition to Dwell Time (per point) and External Trigger delay if enabled.

Fast Sweep    Available ONLY with Opt. 103  When checked, in Analog Sweep mode the analyzer source settling times are shortened in both frequency and power-control (ALC) circuits. In Stepped Sweep mode, the settling time at ALL data points are shortened. This nearly doubles the sweep speed at preset conditions, but at the expense of frequency accuracy and a few dB of amplitude variation. For ratioed measurements, such as S-Parameters, these errors substantially ratio out.

  • By default, Fast Sweep is always OFF to provide maximum accuracy and stability.

  • Fast Sweep is NOT allowed with Power Limit enabled.

  • Note: Performance specifications do NOT apply in Fast Sweep.

Sweep Sequence

Standard Sweep  When checked, the analyzer sweeps all data points for each source port in turn. For a 2-port analyzer, this means that all data points are swept in the forward direction, then all data points are swept in the reverse direction.  Even when NO reverse parameters are displayed (S22 or S12), reverse measurements are necessary when a full 2-port calibration is correcting the channel. This is the default behavior.

Point Sweep  Available ONLY on standard S-parameter channels.  When checked, the analyzer measures all parameters at each frequency point before stepping to the next frequency. The display trace is updated as each data point is measured.

  • Point sweep usually results in slower sweeps and is useful only in rare circumstances.

  • Point sweep is the same as stepped sweep mode on the 8510 and 8530.