How to Configure for GPIB, SCPI, and SICL

M9370A / M9371A / M9372A / M9374A / M9375A

Configure for GPIB, SCPI, and SICL


The following settings are used to configure the analyzer for remote control using SCPI commands.

How to Configure for SICL / GPIB Operation

Using front-panel hardkey [softkey] buttons

Using Menus

  1. Press System

  2. then [Configure]

  3. then [SICL/GPIB]

  1. Click Utility

  2. then System

  3. then Configure

  4. then SICL/GPIB/SCPI

 

SICL / GPIB dialog box help

 

HiSLIP

HiSLIP is available ONLY when the soft front panel has already been launched.

When checked, the PXI VNA will also advertise itself using port TCPIP port 4880. You must configure your firewall to enable communication through this port. You can connect to the module and send SCPI commands by using the VISA address:
TCPIP0::<PXI controller name >::hislip0::INSTR.  ("hislip" is case-sensitive)

For example, if your PXI controller is named: myPXI, then the TCPIP connection string would be:

TCPIP0::myPXI::hislip0::INSTR

When configuring an embedded controller, or when configuring an external computer that is connected using a PXIe cable, then replace <PXI controller name> with "Localhost". For example:

TCPIP0::Localhost::hislip0::INSTR

SCPI Monitor / Input

GPIB Command Processor Console  Launches a window that is used to send single SCPI/GPIB commands from the analyzer keyboard.

Note: Press Control+Z , then enter, to close the console window.

Note: The Status Register system can NOT be used from the GPIB Console.

  • Type a valid command, with appropriate arguments and press enter.

  • Use the arrow keys to recall previous commands.

  • The console window may launch behind the analyzer application. Press Control+Tab to bring the console window to the top.

Local and Remote Operation

The analyzer LCL and  RMT (Local and Remote) operation labels appear in the lower right corner of the status bar.

Note: The status bar is NOT visible when the analyzer is preset. See how to make the status bar visible.

  • LCL appears when NOT under SCPI control

  • RMT appears when under SCPI control. The RMT label does NOT appear when under COM control. Remote operation disables the front panel keys except for the Macro/Local key.

To return to Local (front panel) operation, press the Macro / Local key

Sending the GPIB "GTL" (go to local) command also returns the analyzer to Local operation.

Sending the GPIB "LLO" (local lockout) command disables the front panel Local button.