SYSTem:CAPability Commands
Reads various capabilities of the analyzer.
SYSTem:CAPability: ALC:POWer: HARDware: | ATTenuator:RECeiver: | EXISts? | MAXimum? | ATTenuator:SOURce: | MAXimum? | DC:RECeiver | DC:SOURce | PORTs: | CATalog? | COUNt? | INTernal | CATalog? | COUNt? | PNUMber? | SOURce | CATalog? | COUNt? | INTernal | CATalog? | COUNt? | RECeiver: | INTernal | COUNt? NBW: PRESet:FREQuency: | MAXimum? | MINimum? WINDows
|
Click on a red keyword to view the command details.
See Also
SYSTem:CAPability:ALC:POWer:MAXimum[:LEVel]? <srcNum>
(Read-only) Returns the maximum leveled source power setting in dB. Learn more about leveled source power. |
|
Parameters |
|
<srcNum> |
Source Number. Choose from 1 or 2. |
Examples |
SYST:CAP:ALC:POW:MAX? 1 |
Return Type |
Numeric |
Not Applicable |
SYSTem:CAPability:ALC:POWer:MINimum[:LEVel]? <srcNum>
(Read-only) Returns the minimum leveled source power setting in dB with 0 dB attenuation. Learn more about leveled source power. |
|
Parameters |
|
<srcNum> |
Source Number. Choose from 1 or 2. |
Examples |
SYST:CAP:ALC:POW:MIN? 1 |
Return Type |
Numeric |
Not Applicable |
SYSTem:CAPability:CHANnels:MAXimum[:COUNt]?
(Read-only) Returns the maximum possible number of channels. Learn more about Channels. |
|
Parameters |
None |
Examples |
SYST:CAP:CHAN:MAX? |
Return Type |
Numeric |
Not Applicable |
SYSTem:CAPability:FOM:EXISts?
(Read-only) Returns whether or not the analyzer has FOM Opt. 080 installed. Learn more. |
|
Parameters |
None |
Examples |
SYST:CAP:FOM:EXIS? |
Return Type |
Boolean 1 - Yes, FOM is installed. 0 - No, FOM is NOT installed. |
Not Applicable |
SYSTem:CAPability:HARDware:ATTenuator:RECeiver:EXISts? <portNum>
(Read-only) Returns whether or not there is a receiver attenuator on the specified port. |
|
Parameters |
|
<portNum> |
Port number. Choose from the number of test ports on the analyzer. |
Examples |
SYST:CAP:HARD:ATT:REC:EXIS? 2 |
Return Type |
Boolean 1 - Yes, the test port has a receiver attenuator. 0 - No, the test port does NOT have a receiver attenuator. |
Not Applicable |
SYSTem:CAPability:HARDware:ATTenuator:RECeiver:MAXimum? <portNum>
(Read-only) Returns the maximum amount of receiver attenuation on the specified port. |
|
Parameters |
|
<portNum> |
Port number. Choose from the number of test ports on the analyzer. |
Examples |
SYST:CAP:HARD:ATT:REC:MAX? 2 |
Return Type |
Numeric |
Not Applicable |
SYSTem:CAPability:HARDware:ATTenuator:RECeiver:STEP[:SIZE]? <portNum>
(Read-only) Returns the step size of the receiver attenuator on the specified port. |
|
Parameters |
|
<portNum> |
Port number. Choose from the number of test ports on the analyzer. |
Examples |
SYST:CAP:HARD:ATT:REC:STEP? 2 |
Return Type |
Numeric |
Not Applicable |
SYSTem:CAPability:HARDware:ATTenuator:SOURce:MAXimum? <portNum>
(Read-only) Returns the maximum amount of source attenuation on the specified port. |
|
Parameters |
|
<portNum> |
Port number. Choose from the number of test ports on the analyzer. |
Examples |
SYST:CAP:HARD:ATT:SOUR:MAX? 2 |
Return Type |
Numeric |
Not Applicable |
SYSTem:CAPability:HARDware:ATTenuator:SOURce:STEP[:SIZE]? <portNum>
(Read-only) Returns the step size of the source attenuator on the specified port. |
|
Parameters |
|
<portNum> |
Port number. Choose from the number of test ports on the analyzer. |
Examples |
SYST:CAP:HARD:ATT:SOUR:STEP? 2 |
Return Type |
Numeric |
Not Applicable |
SYSTem:CAPability:HARDware:DC:RECeiver:INTernal:CATalog?
(Read-only) Returns a list of names of the internal DC receivers. |
|
Parameters |
None |
Examples |
SYST:CAP:HARD:DC:REC:INT:CAT? |
Return Type |
String of internal DC receivers separated by commas. For example, "AI1,AI2,AIG,AOS1,AOS2" |
Not Applicable |
SYSTem:CAPability:HARDware:DC:RECeiver:INTernal:COUNt?
(Read-only) Returns the number of internal DC receivers in the analyzer. |
|
Parameters |
None |
Examples |
SYST:CAP:HARD:DC:REC:INT:COUN? |
Return Type |
Numeric |
Not Applicable |
SYSTem:CAPability:HARDware:DC:SOURce:INTernal:CATalog?
(Read-only) Returns a list of names of the internal DC sources. |
|
Parameters |
None |
Examples |
SYST:CAP:HARD:DC:SOUR:INT:CAT? |
Return Type |
String of internal DC sources separated by commas. For example, "AO1,AO2" |
Not Applicable |
SYSTem:CAPability:HARDware:DC:SOURce:INTernal:COUNt?
(Read-only) Returns the number of internal DC sources in the analyzer. |
|
Parameters |
None |
Examples |
SYST:CAP:HARD:DC:SOUR:INT:COUN? |
Return Type |
Numeric |
Not Applicable |
SYSTem:CAPability:HARDware:PORTs:CATalog?
(Read-only) Returns a list of test port names including external testset ports. |
|
Parameters |
None |
Examples |
SYST:CAP:HARD:PORT:CAT? |
Return Type |
String of port names separated by commas. For example, "Port 1,Port 2,Port 3,Port 4". |
Not Applicable |
SYSTem:CAPability:HARDware:PORTs:COUNt?
(Read-only) Returns the number of test ports including external testset ports. |
|
Parameters |
None |
Examples |
SYST:CAP:HARD:PORT:COUN? |
Return Type |
Numeric |
Not Applicable |
SYSTem:CAPability:HARDware:PORTs:INTernal:CATalog?
(Read-only) Returns a list of internal test port names. |
|
Parameters |
None |
Examples |
SYST:CAP:HARD:PORT:INT:CAT? |
Return Type |
String of port names separated by commas. For example, "Port 1,Port 2,Port 3,Port 4". |
Not Applicable |
SYSTem:CAPability:HARDware:PORTs:INTernal:COUNt?
(Read-only) Returns the number of internal test ports. |
|
Parameters |
None |
Examples |
SYST:CAP:HARD:PORT:INT:COUN? |
Return Type |
Numeric |
Not Applicable |
SYSTem:CAPability:HARDware:PORTs:PNUMber? <portName>
(Read-only) Returns the port number associated with the specified port name. |
|
Parameters |
None |
<portName> |
String. Port name. Use SYST:CAP:HARD:PORT:CAT? to return a list of valid port names. |
Examples |
SYST:CAP:HARD:PORT:PNUM? "Port 1" |
Return Type |
Numeric |
Not Applicable |
SYSTem:CAPability:HARDware:PORTs:SOURce:CATalog?
(Read-only) Returns a list of source port names, including any configured external sources. |
|
Parameters |
None |
Examples |
SYST:CAP:HARD:PORT:SOUR:CAT? |
Return Type |
String of source port names separated by commas. For example, "Port 1,Port 2,Port 3,Port 4, Port 1 Src 2". |
Not Applicable |
SYSTem:CAPability:HARDware:PORTs:SOURce:COUNt?
(Read-only) Returns the number of source ports, including any configured external sources. |
|
Parameters |
None |
Examples |
SYST:CAP:HARD:PORT:SOUR:COUN? |
Return Type |
Numeric |
Not Applicable |
SYSTem:CAPability:HARDware:PORTs:SOURce:INTernal:CATalog?
(Read-only) Returns a list of internal source port names. |
|
Parameters |
None |
Examples |
SYST:CAP:HARD:PORT:SOUR:INT:CAT? |
Return Type |
String of internal source port names separated by commas. For example, "Port 1,Port 2,Port 3,Port 4, Port 1 Src 2" |
Not Applicable |
SYSTem:CAPability:HARDware:PORTs:SOURce:INTernal:COUNt?
(Read-only) Returns the number of internal source ports. |
|
Parameters |
None |
Examples |
SYST:CAP:HARD:PORT:SOUR:INT:COUN? |
Return Type |
Numeric |
Not Applicable |
SYSTem:CAPability:HARDware:RBSWitch:EXISts? <portNum>
(Read-only) Returns whether or not the specified port number has a reference bypass switch. |
|
Parameters |
|
<portNum> |
Port number. Choose from the number of test ports on the analyzer. |
Examples |
SYST:CAP:HARD:RBSW:EXIS? 2 |
Return Type |
Boolean 1 - Yes, the test port has a reference bypass switch. 0 - No, the test port does NOT have a reference bypass switch. |
Not Applicable |
SYSTem:CAPability:HARDware:RECeiver:INTernal:COUNt?
(Read-only) Returns the number of receivers in the analyzer. |
|
Parameters |
None |
Examples |
SYST:CAP:HARD:REC:INT:COUN? |
Return Type |
Numeric |
Not Applicable |
SYSTem:CAPability:HARDware:SOURce:COUNt?
(Read-only) Returns the number of sources in the analyzer. |
|
Parameters |
None |
Examples |
SYST:CAP:HARD:SOUR:COUN? |
Return Type |
Numeric |
Not Applicable |
SYSTem:CAPability:NBW:NOISe:CATalog?
(Read-only) Returns the list of supported Noise Bandwidth values when using a noise receiver (option 029). Learn more about Opt. 029. |
|
Parameters |
None |
Examples |
SYST:CAP:NBW:NOIS:CAT? |
Return Type |
Variant array of string values |
Not Applicable |
SYSTem:CAPability:NBW:STD:CATalog?
(Read-only) Returns the list of supported Noise Bandwidth values when using the NA receiver for noise measurements (option 028). Learn more about Opt 028. |
|
Parameters |
None |
Examples |
SYST:CAP:NBW:STD:CAT? |
Return Type |
Variant array of string values |
Not Applicable |
SYSTem:CAPability:PRESet:FREQuency:MAXimum?
(Read-only) Returns the maximum specified frequency of the analyzer. Does not include any over-sweep. See also: SYST:CAP:FREQ:MAX? |
|
Parameters |
None |
Examples |
SYST:CAP:PRES:FREQ:MAX? |
Return Type |
Numeric |
Not Applicable |
SYSTem:CAPability:PRESet:FREQuency:MINimum?
(Read-only) Returns the minimum specified frequency of the analyzer. Does not include any under-sweep. See also: SYST:CAP:FREQ:MIN? |
|
Parameters |
None |
Examples |
SYST:CAP:PRES:FREQ:MIN? |
Return Type |
Numeric |
Not Applicable |
SYSTem:CAPability:RBW:IMS:CATalog?
(Read-only) Returns the list of supported Resolution BW values for the IMSpectrum channel. Learn more about IMSpectrum. |
|
Parameters |
None |
Examples |
SYST:CAP:RBW:IMS:CAT? |
Return Type |
Variant array of string values |
Not Applicable |
SYSTem:CAPability:WINDows:MAXimum[:COUNt]?
(Read-only) Returns the maximum number of windows. Learn more. |
|
Parameters |
None |
Examples |
SYST:CAP:WIND:MAX? |
Return Type |
Numeric |
Not Applicable |
SYSTem:CAPability:WINDows:TRACes:MAXimum[:COUNt]?
(Read-only) Returns the maximum number of traces per window. Learn more. |
|
Parameters |
None |
Examples |
SYST:CAP:WIND:TRAC:MAX? |
Return Type |
Numeric |
Not Applicable |