Power Level

M9370A / M9371A / M9372A / M9374A / M9375A

Power Level


Power level is the power of the source at the test ports.

See other 'Setup Measurements' topics

Power Settings

The test port output power is specified over frequency.

See the Power Range specifications for your analyzer.

How to make Power settings

Use one of the following methods to set port power.

Using front-panel hardkey [softkey] buttons

Using Menus

  1. Press Power

  2. then [Power] or [Power and Attenuators]

  1. Click Stimulus

  2. then Power or Power and Attenuators

 

Power dialog box help

This dialog provides basic control of source power for a specific port.

See Power and Attenuators dialog box.

Power On (All Channels)  Check to enable source power for all channels. Only turns power ON if channel power setting is ON or Auto.

Port 'n'  Active source port for which power is being set.

Port Power  Sets the power level for the specified port.

Power Sweep

Start / Stop Power  Set the start and stop power values of a power sweep.

  • These settings are only available when Sweep Type is set to Power Sweep.

  • You can Zoom to easily change the start and stop power levels in a power sweep. Learn how.

Power Slope

Not available on M937x models.

 

Power and Attenuators dialog box help

Defines and controls the source power and attenuation for the active channel.

External sources can be controlled from this dialog. Learn more.

Power On (All Channels)  Check to enable source power for all channels. Only turns power ON if channel power setting is ON or Auto.

Port Powers Coupled

  • Coupled (checked)  The power levels are the same at each test port. Set power at any test port and all test ports change to the same power level.

  • Uncoupled (cleared)  The power levels are set independently for each test port. Uncouple power, for example, if you want to measure the gain and reverse-isolation of a high-gain amplifier. The power required for the input port of the amplifier is much lower than the power required for the output port. A power sweep can also be performed with uncoupled power.

Name  Lists the analyzer test ports.

State

  • Auto  Source power is turned ON at the specified test port when required by the measurement. This is the most common (default) setting. See also Power ON and OFF during Save / Recall, User Preset, and Preset.

  • ON  Source power is ALWAYS ON, regardless of measurements that are in process. Use this setting to supply source power to a DUT port that always requires power, such as an LO port. This could turn OFF power at another test port. Learn about internal second source restrictions.

  • OFF  Source power is never ON, regardless of the measurement requirements. Use this setting to prevent damage to a sensitive DUT test port.

Port Power  Sets the power level at the output of the source.

Start / Stop Power  Available ONLY when sweep type is set to Power Sweep. Set the start and stop power values of a power sweep. Learn how to set Power Sweep.

  • You can specify whether to maintain source power at either the start power or stop power level at the end of a power sweep. Learn more.

  • A power sweep can be performed with uncoupled power. Different power ranges can be swept in the forward and reverse directions.

Leveling Mode

  • Internal - Source correction factors are used to provide a flat output power within specifications.  NO internal circuitry is used to level the output power. Therefore, the source will NEVER become unleveled.

  • Receiver Rx - Receiver Leveling. Select a receiver to use for leveling the source. Learn more.

Channel Power Slope

Not available on M937x models.

Offset and Limits  Launches the Power Offset and Limits dialog.

Power ON and OFF during Save / Recall, User Preset, and Preset

To protect your DUT from being inadvertently powered ON, the following RF Power ON/OFF settings occur:

Instrument State Save/Recall

If power is OFF when an instrument state is saved, then power will always be OFF after the instrument state is recalled.

If power is ON when an instrument state is saved, and the current power setting is OFF, then power will be OFF after the instrument state is recalled.

Preset

Instrument Preset sets power ON by default.

This can be changed with a Preference setting so that, if the current power setting is OFF, then power will be OFF after Preset.

Power ON and OFF during Sweep and Retrace

Source power is NOT turned OFF during M937xA frequency band crossings or during sweep retrace.

Caution: Avoid expensive repairs to your analyzer. Read Electrostatic Discharge Protection.