Power Level
Power level is the power of the source at the test ports.
See other 'Setup Measurements' topics
The test port output power is specified over frequency.
See the Power Range specifications for your analyzer.
How to make Power settingsUse one of the following methods to set port power. |
|
Using front-panel hardkey [softkey] buttons |
Using Menus |
|
|
This dialog provides basic control of source power for a specific port. See Power and Attenuators dialog box. Power On (All Channels) Check to enable source power for all channels. Only turns power ON if channel power setting is ON or Auto. Port 'n' Active source port for which power is being set. Port Power Sets the power level for the specified port. Power SweepStart / Stop Power Set the start and stop power values of a power sweep.
Power SlopeNot available on M937x models. |
Defines and controls the source power and attenuation for the active channel. External sources can be controlled from this dialog. Learn more. Power On (All Channels) Check to enable source power for all channels. Only turns power ON if channel power setting is ON or Auto.
Name Lists the analyzer test ports. State
Port Power Sets the power level at the output of the source. Start / Stop Power Available ONLY when sweep type is set to Power Sweep. Set the start and stop power values of a power sweep. Learn how to set Power Sweep.
Channel Power SlopeNot available on M937x models. Offset and Limits Launches the Power Offset and Limits dialog. |
Power ON and OFF during Save / Recall, User Preset, and Preset
To protect your DUT from being inadvertently powered ON, the following RF Power ON/OFF settings occur:
Instrument State Save/Recall
If power is OFF when an instrument state is saved, then power will always be OFF after the instrument state is recalled.
If power is ON when an instrument state is saved, and the current power setting is OFF, then power will be OFF after the instrument state is recalled.
Preset
Instrument Preset sets power ON by default.
This can be changed with a Preference setting so that, if the current power setting is OFF, then power will be OFF after Preset.
Power ON and OFF during Sweep and Retrace
Source power is NOT turned OFF during M937xA frequency band crossings or during sweep retrace.
Caution: Avoid expensive repairs to your analyzer. Read Electrostatic Discharge Protection.