Load Eterms into Cal Sequence

M9370A / M9371A / M9372A / M9374A / M9375A

Load Error Terms during a Cal Sequence


This example requires that you already have a Cal Set named "foo" that contains a 1-port cal on port 1 and a 1-port cal on port 2.

This example starts a Guided Calibration specifying an Unknown Thru. It loads the 1-port Cals from the existing "foo" Cal Set, then recalculates the number of steps required to complete the cal. After loading the 1-port cals, only the Unknown Thru standard is left to acquire.

SENS:CORR:COLL:GUID:CONN:PORT1 "APC 3.5 female"

SENS:CORR:COLL:GUID:CONN:PORT2 "APC 3.5 female"

SENS:CORR:COLL:GUID:CKIT:PORT1 "85033D/E"

SENS:CORR:COLL:GUID:CKIT:PORT2 "85033D/E"

SENS:CORR:COLL:GUID:METH UNKN

' auto-create user calsets for SCPI

SENS:CORR:PREF:CSET:SAVU 1

SENS:CORR:COLL:GUID:INIT

' should return the number 7

SENS:CORR:COLL:GUID:STEPS?

' to port 1, from port 1 in calset

SENS:CORR:COLL:GUID:ETER:LOAD "foo",1,1

' to port 2, from port 2 in calset

SENS:CORR:COLL:GUID:ETER:LOAD "foo",2,2

' should now return the number 1

SENS:CORR:COLL:GUID:STEPS?

' measure the unknown thru

SENS:CORR:COLL:GUID:ACQ STAN1

' save the cal to new user calset

SENS:CORR:COLL:GUID:SAVE