PXIFPTour

M9370A / M9371A / M9372A / M9374A / M9375A

N937xA Front Panel Tour


See Also


Status LED

The following LED colors indicate the M937xA status:

Green (solid) – Firmware is running but the M937xA is NOT sweeping.

Green (blinking) - Each blink indicates a measurement sweep. The M937xA may be sweeping faster than the indicator can blink.

Amber – Power is ON but the firmware is not running.

Red – A hardware error has occurred.

Off – Power is OFF.

Test Ports

All M937xA models are available with 2 test ports.

See Specs for more information about the Test port connectors and Input damage levels.

Ref In / OUT

IN  When a 10 MHz external reference signal is detected at this port, it will be used as the instrument frequency reference instead of the internal frequency reference.

OUT  This SMB(m) connector outputs a 10 MHz frequency reference signal for use by other modules and test equipment.

Note: To use an external reference signal, you must make a setting in the M937xA.

  • From the soft front panel, click Utility, then System, then Configure, then check External Reference.

  • From SCPI, use SENS:ROSC:SOUR

Trigger Lines

Trig IN  When enabled, the module accepts signals on this connector which indicates that the external devices is ready to be triggered.

Trig OUT  When enabled, the module outputs signals on these connectors either before or after a measurement.

Trig Ready  When enabled, the module outputs a 'READY' signal on this connector to other devices. This indicates that the module is ready to be triggered.

Learn more about External Triggering

LO Out / LO In

For Daisy-chaining two or more modules. See the M937xA Installation Guide for more information.