NI PXI-2565 Relay Replacement
The NI PXI-2565 uses socketed electromechanical relays.
Refer to the following tables for information about ordering replacement relays.
Relay Manufacturer | Part Number |
---|---|
Aromat (NAiS) | DSP1a-DC5V |
Relay Kit | Part Number |
---|---|
National Instruments (16 relays) | 777880-01 |
Complete the following sets of steps to disassemble your module, replace a failed relay, and reassemble your module.
Disassemble the Module
- Ground yourself using a grounding strap or a ground connected to your PXI chassis.
Note Properly grounding yourself prevents damage to your module from electrostatic discharge. - Remove the four screws that fasten the NI PXI-2565 relay module to the switch carrier module and front panel.
- Lift up the card as shown in the figure below.
1 Carrier Module 2 Solder Location for
Flyback Voltage Protection Device3 Relay Module 4 Screws 5 Insulator 6 Relay and Socket 7 Ribbon Cable - Remove the relay by gently applying force.
- Before inserting the new relay, match the direction of the relay and the socket.
- Insert the relay, making sure the relay is properly seated and the socket hooks the top of the relay.
- Reassemble your module.