NI PXI-2529 2-Wire 4×32 Matrix Topology

NI Switches

NI PXI-2529 2-Wire 4×32 Matrix Topology

The NI TB-2634 or the NI TB-2636 terminal block creates a 2-wire 4×32 matrix topology with the NI PXI-2529. The following figure represents the NI PXI-2529 in the 2-wire 4×32 matrix topology.

Making a Connection

The NI PXI-2529 combined with the NI TB-2634 or the NI TB-2636 creates a 4×32 matrix. In this topology, you can connect any row to any column. When connecting signals to r1, you would connect them to R1+ and R1– in the terminal block. When connecting signals to c0, you would connect them to C0+ and C0– in the terminal block.

You can connect the channels of the NI PXI-2529 using the niSwitch Connect Channels VI or the niSwitch_Connect function. For example, to connect row 1 to column 1, call the niSwitch Connect Channels VI or the niSwitch_Connect function with the channel 1 parameter set to r1 and the channel 2 parameter set to c1.

When scanning the NI PXI-2529 in 4×32 matrix topology, a typical scan list entry could be r1->c20;. This entry routes the signal connected to row 1 to column 20.

Bank Connection Diagram

Terminal Block Connections

The NI TB-2634 terminal block creates the following connections or relabels pins to operate the NI PXI-2529 as a 2-wire 4×32 matrix topology. Refer to the following pinout for pin locations.

The following table lists the pins of the NI TB-2634 shorts and the column connection created.

ConnectionPinsConnectionPinsConnectionPinsConnectionPins
C0+B24C8+B19C16+B14C24+B9
C0-A24C8-A19C16-A14C24-A9
C1+B23C9+B18C17+B13C25+B8
C1-A23C9-A18C17-A13C25-A8
C2+B22C10+B17C18+B12C26+B7
C2-A22C10-A17C18-A12C26-A7
C3+B21C11+B16C19+B11C27+B6
C3+A21C11-A16C19-A11C27-A6
C4+C24C12+C19C20+C14C28+C9
C4-D24C12-D19C20-D14C28-D9
C5+C23C13+C18C21+C13C29+C8
C5-D23C13-D18C21-D13C29-D8
C6+C22C14+C17C22+C12C30+C7
C6-D22C14-D17C22-D12C30-D7
C7+C21C15+C16C23+C11C31+C6
C7-D21C15-D16C23-D11C32-D6

The following table lists the pins of the NI TB-2634 shorts and the row connection created.

ConnectionPinsConnectionPinsConnectionPinsConnectionPins
R0+B25,B15R1+C25,C15R2+B20,B10R3+C20,C10
R0-A25,A15R1-D25,D15R2-A20,A10R3-D20,D10

Pinout

The following figure identifies the pins for the NI PXI-2529.