NI SCXI-1193 16×1 Terminated Multiplexer Topology
The following figure represents the NI SCXI-1193 16×1 terminated multiplexer topology.
The following figure shows the reset position of the NI SCXI-1193 in the 16×1 terminated multiplexer topology.
Making a Connection
For proper termination, connect an external terminator, such as the NI 50 Ω MCX terminator (778831-01), to every odd channel. Any input connected to an even channel not connected to the COM is automatically routed to its associated termination channel.
The NI SCXI-1193 in this topology contains 16 input channels connected to a common channel. These input channels are the even channels from channel 0 to channel 30. The common channel is referred to as com0. You can connect any even input channel to com0 in this topology.
You can connect the channels of the NI SCXI-1193 using the niSwitch Connect Channels VI or the niSwitch_Connect function. For example, to connect channel 16 to common 0, call the niSwitch Connect Channels VI or the niSwitch_Connect function with the channel 1 parameter set to ch16 and the channel 2 parameter set to com0.
When scanning the NI SCXI-1193, a typical scan list entry could be ch2->com0;. This entry disconnects ch2 from its termination and routes it to com0.