NI signal generators can correct for slight deviations in the flatness of the frequency characteristic of the analog lowpass filter in its passband, as shown in the following figure.

Curve A shows a typical lowpass filter curve. The response of the filter is stored in an onboard EEPROM in 1 MHz increments up to:

  • 16 MHz for NI PXI/PCI-5401/5411
  • 8 MHz for NI PXI/PCI-5431

Curve C is the correction applied to the frequency response. The resulting curve B is a flat response over the entire passband. If you want to generate a sine wave at a particular frequency with filter correction applied, you must specify that frequency through the software.

Note  You can change the filter frequency correction at any time during waveform generation.