Save Custom Analysis Format #6 (H.W.)

EPMA Probe

 

Probe Menu Details > Output

Save Custom Analysis Format #6 (H.W.)

This menu (H.W. format) will create an output file of standard or unknown atomic, elemental and oxide weight percents, along with single point detection and analytical sensitivity and raw k-ratios, count times, beam current and x, y, z coordinates and relative distance in microns.

 

            Sample description text (no column label)

LINE        Line number

El AT%      Atomic percent

El WT%      Elemental weight percent

El (O2)     Oxide weight percent

TOTAL       Total weight percent

El CDL99    Detection limit in elemental weight percent at 99% confidence

El %ERR     Analytical error in relative percent (one sigma)

El K-RAW    Elemental raw k-ratio (unk counts/standard counts) (includes x-ray line, spectro number and crystal)

El ONTIM    On-peak counting time(includes x-ray line, spectro number and crystal)

El HITIM    High off-peak counting time(includes x-ray line, spectro number and crystal)

El LOTIM    Low off-peak counting time(includes x-ray line, spectro number and crystal)

BEAMCUR     Beam current (in nA)

X-POS       X stage position

Y-POS       Y stage position

Z-POS       Z stage position

RELDIST     Relative distance (in microns)

 

This file format was specified by Heather Wright at the University of Oregon and cannot be modified for other applications.  In addition the first column of tab delimited data will contain any sample description text as separate strings enclosed in double quotes. This is the only output format in PROBE that will export the sample description field text.