Peak Center Options
· Acquire Automated PHA Scan Prior To Peaking
· Acquire Automated PHA Scan After Peaking
These two options will allow the program to automatically acquire a PHA scan for each element selected, either before, after or both before and after each peak centering. The results of the PHA scan are automatically saved to the user’s probe database and can be displayed using the Run | Display and Export Scans menu.
· Display PHA Dialog Prior To Peaking
· Display PHA Dialog After Peaking
These two options will allow the program to display the manual PHA dialog for each element selected, either before, after or both before and after each peak centering. All three PHA scan types can be acquired: PHA, bias and gain scans. The results of the PHA scan are automatically saved to the user’s probe database and can be displayed using the Run | Display and Export Scans menu.