Low-Current Measurement Considerations
You can use the NI 4070/4071/4072 to measure low current levels.
The NI 4071 offers more current measurement ranges than the NI 4070/4072. In the 1 µA range of the NI 4071, you can resolve currents down to picoamperes.
Low-level measurements are susceptible to sources of error and noise that are often negligible for higher level measurements. In the low-level signals world, no material is an ideal insulator. Some materials are better insulators than others. Selection of proper cables and interconnects becomes very important.
When making low-level current measurements, noise and error sources are more likely to affect measurements in the order of 10µA or less. Take into account the recommendations listed in the table below to ensure measurement integrity.
Error or Noise Source | Cause | Level of Induced Error | Recommendations |
---|---|---|---|
Input bias current |
|
Typically 10 pA at 23 °C ambient, but it is factory calibrated to zero at that temperature. |
|
Leakage currents due to poor insulation | When measuring low-level currents, leakage currents in insulators become relevant. Current flows between a high-impedance point and the nearby voltage sources. | Depends upon the resistivity of the insulator material. Some examples include the following:
|
Be sure that the test fixture for the DUT is constructed of a material appropriate for the desired test and that the test fixture is clean. |
Noise currents due to triboelectrics1 | Currents are generated by moving the cables because a transfer of charge occurs between the insulator and the conductor when they rub against each other. | Hundreds of pA in a polyethylene cable (for example, RG-58) |
|
Noise currents due to piezoelectrics2 | Currents can be generated when the insulator is subjected to mechanical stress. | Hundreds of pA for polyethylene |
|
Error currents due to contamination and humidity | Insulation resistance gets reduced with increasing humidity and contamination. These two combined could also generate small electrochemically-induced currents. | On the order of pA to µA |
|
Noise currents due to electromagnetic interference and electric fields | Surrounding equipment (motors, power cables, vibrating equipment) can induce noise currents. Moving people and charged objects near measurement setup can induce electrostatic coupling. | On the order of nA and tens of uA depending on proximity, device shielding, and magnitude of relative movement |
|
Burden voltage | For the maximum burden voltage per range, refer to the Related Documentation for the NI 4070/4071/4072 specifications documents. | Depends on the circuit being measured | Calculate if the error caused by burden voltage is relevant when measuring the current in the circuit of interest. Refer to Burden Voltage for an example. |
Leakage currents due to external switching | External switches used for routing low level currents could be a source of leakage currents. | Depends on the switch, temperature and humidity of the environment. |
|
1The Triboelectric Effect can be compared to the phenomenon of static electricity caused by rubbing certain insulators together. Teflon and silver in combination, for example, can create a very high level of triboelectric-induced current noise. 2The Piezoelectric Effect, although well-exploited in sensors which convert changes in pressure from sound waves or physical vibration into small voltage signals, is highly undesirable in cables and interconnects. |
In addition to following these recommendations to reduce noise and errors, use software filtering and noise reduction techniques such as increasing the measurement aperture, choosing the most appropriate type of DC Noise Rejection, or averaging several measurements by setting the Number of Averages property.
Refer to Switching Current for more recommendations for systems with switches.