Configuring Period Measurements
For the NI 4070/4071/4072 only, configure a period measurement by completing the following steps:
Note If you specify Auto Range (-1.0) for the period range, the DMM defaults to a 2 second aperture time. |
- LabVIEW
- Use niDMM Config Measurement.
- Set Measurement Function to Period.
- Set Range to the maximum expected period in seconds. The DMM uses the range to calculate the aperture time.
- Use niDMM Configure Frequency Voltage Range.
- Set Frequency Voltage Range to the AC V range of your input signal.
- CVI/C++/Visual Basic
- Use niDMM_ConfigureMeasurementDigits.
- Set Measurement Function to NIDMM_VAL_PERIOD.
- Set Range to the maximum expected period in seconds. The DMM uses the range to calculate the aperture time.
- Use niDMM_ConfigureFrequencyVoltageRange
- Set Frequency Voltage Range to the AC V range of your input signal.