DMM Measurement Defaults
The following table lists the default measurement settings for the NI 4070/4071/4072.
Function | Aperture | Auto Zero | ADC Calibration | DC Noise Rejection |
---|---|---|---|---|
DC 7½ digits1 | 100 ms | ON | ON | High-Order |
DC 6½ digits | 100 ms | ON | ON | High-Order |
DC 5½ digits |
| ONCE | OFF | Second-Order |
DC 4½ digits |
| ONCE | OFF | Second-Order |
DC Auto Range | 500 µs | ONCE | OFF | Second-Order |
AC | Max (DC or 4/(minFreq)) | OFF | OFF | N/A |
AC Auto Range | Max (500 µs or 4/(minFreq)) | OFF | OFF | N/A |
Frequency | 2/(minFreq) | N/A | N/A | N/A |
Period | 2 x (the maximum period) | N/A | N/A | N/A |
Capacitance and Inductance |
|
N/A | N/A | N/A |
1The default Number of Averages for the NI 4071 is 4. 2For the NI 4071, on the 5.5 DCV 100V range, use 1.6 ms. |
By default, the Aperture time, Auto Zero, ADC Calibration, DC Noise Rejection settings for a measurement are chosen by the driver based on the configured measurement and resolution. These values are chosen to ensure accuracy for 6½–digit or 7½–digit measurements while not sacrificing performance at lower resolutions. For more information on a particular attribute, including how to configure a non-default setting, refer to Features. For AC, excluding Auto Range, the aperture is expressed as Max (DC or 4/minFreq). minFreq, or minimum frequency, is a user–programmable parameter with a default value of 20 Hz. Therefore, the aperture default is 200 ms.
The same applies for other user selectable values of minFreq. For example, on the NI 4070/4071/4072, if a minFreq of 1 kHz is selected then 4/minFreq = 4/1 kHz = 4 ms. In this case 4 ms would be used as the AC aperture for resolutions <6 digits. For resolutions >6 digits, 100 ms would be used.
To adjust the default aperture time, refer to Configuring Measurement Timing.
Default Settling Times
Function | Settling Time |
---|---|
DC V (100 mV–10 V) | 1 ms |
DC V (100 V, 300 V) (NI 4070/4072 only) | 2 ms |
DC V (100 V, 1000 V) (NI 4071 only) | 5 ms |
DCV Auto Range | 2 ms |
Resistance ≤1 kΩ | 1 ms |
Resistance 10 kΩ | 5 ms |
Resistance 100 kΩ | 25 ms |
Resistance 1 MΩ | 100 ms |
Resistance ≥ 10 MΩ | 250 ms |
Resistance Auto Range | 50 ms |
AC V DC coupled | 3 µs |
AC V AC coupled | 1 s |
Frequency/Period | 500 ms |
DC I | 100 µs |
AC I | 3 µs |
Diode | 10 ms |
Inductance (NI 4072 only) | 3 µs |
Capacitance (NI 4072 only) | 3 µs |
To adjust the default settling times, refer to Configuring Measurement Timing.