Autosetup |
![]() ![]() ![]() |
![]() |
The Autosetup feature in PicoScope sets up timebase, channels and trigger configuration automatically. It supports all converters. Activate it by clicking the "lightning" button.
|
AutoSetup follows these rules:
![]() | The timebase will auto-range to the first enabled channel with a signal.
|
![]() | If no channels have signals then the timebase will be set to the default timebase for that converter.
|
![]() | If there is a trigger enabled, the trigger channel is automatically set to the first enabled channel with a signal.
|
![]() | The trigger mode is set to Auto and the trigger delay to -10%. The trigger threshold is set to 0mV if AC coupling is enabled or is centred between the peak values if DC coupling is enabled.
|
![]() | If there is a trigger enabled and there are no signals in any channels, the trigger is disabled.
|
![]() | The voltage range is set relative the signal on that channel. If no signal is present, it is set to the minimum range for that channel.
|
![]() | The AC/DC coupling state is not changed.
|
![]() | If a channel is enabled, it will stay enabled and if it is disabled it will stay disabled.
|
Autosetup can detect frequencies from 50 Hz to a quarter of the sample rate of the scope.