niScope Cal Fetch Internal Reference
Returns the value of the internal reference that is stored with niScope Cal Store Internal Reference during external calibration. This VI allows tracking of the internal reference value, although the value is not used during digitizer operation or self-calibration. The internal reference value is stored as a 32-bit floating point number in the EEPROM.
instrument handle identifies a particular instrument session. | |||||||
which reference specifies which reference to store, fetch, or route. | |||||||
error in describes error conditions that occur before this VI runs. The default input of this cluster is no error. If an error already occurred, this VI returns the value of error in in error out. The VI runs normally only if no incoming error exists. Otherwise, the VI passes the error in value to error out. The error in cluster contains the following parameters:
| |||||||
instrument handle out has the same value as the instrument handle. | |||||||
internal reference is the value of the internal reference stored during the last external calibration. | |||||||
error out contains error information. If error in indicates that an error occurred before this VI or function ran, error out contains the same error information. Otherwise, it describes the error status that this VI or function produces.
|