Triggering:Onboard Signal Processing:Ref Trig Min Quiet Time
Short Name: Ref Trig Min Quiet Time
Specifies the amount of time (in seconds) the trigger circuit must not detect a signal above the trigger level (or below the trigger level if the trigger slope is negative) before the trigger is armed. This attribute is useful for triggering at the beginning of signal bursts instead of in the middle of signal bursts.
Note This attribute can be used only with high-speed digitizers that support onboard signal processing (OSP). NI-SCOPE returns an error if you use this attribute with a device that does not support OSP. |
Default Value: 0
Valid Values
Any value greater than or equal to 0.
Remarks
The following table lists the characteristics of this property.
Data Type | ViReal64 |
Access | R/W |
Channel Based | No |