SMC-Based Digitizers Property Default Values

NI-SCOPE LabVIEW

SMC-Based Digitizers Property Default Values

The following table lists the property default values for NI SMC-based digitizers. N/A indicates that the attribute is not supported by the device.

Property (Attribute)Default Value
Acq Start Time0
Acquisition TypeNormal
Actual Record Length 1000
Actual Sample Rate 1000000
Advance Trigger SourceNo Source
Arm Reference Trigger SourceNo Source
Array Gain 1
Array Offset 0
Backlog0
Binary Sample Width NI 5105: 16
NI 5114: 8
NI 5122: 16
NI 5124: 16
NI 5142: 16
NI 5152: 8
NI 5922: 32
Center Frequency NI 5105: N/A
NI 5114: N/A
NI 5122: N/A
NI 5124: N/A
NI 5142: 10 MHz
NI 5152: N/A
NI 5922: N/A
Channel Based High Ref 90
Channel Based Low Ref 10
Channel Based Mid Ref 50
Channel Count NI 5105: 8
NI 5114: 2
NI 5122: 2
NI 5124: 2
NI 5142: 2
NI 5152: 2
NI 5922: 2
Channel Enabled 0
Class Group Capabilities IviScopeBase, IviScopeWaveformMeasurement, IviScopeSampleMode, IviScopeAutoSetup, IviScopeTVTrigger
Clock Sync Pulse Source No Source
Combined Decimation N/A
Data Processing Mode NI 5105: N/A
NI 5114: N/A
NI 5122: N/A
NI 5124: N/A
NI 5142: Complex
NI 5152: N/A
NI 5922: N/A
DDC Enabled NI 5105: N/A
NI 5114: N/A
NI 5122: N/A
NI 5124: N/A
NI 5142: FALSE
NI 5152: N/A
NI 5922: N/A
Delay Before Initiate 0
Description National Instruments High Speed Digitizers Oscilloscope Instrument Driver
Digital Gain NI 5105: N/A
NI 5114: N/A
NI 5122: N/A
NI 5124: N/A
NI 5142: 1
NI 5152: N/A
NI 5922: N/A
Digital Offset NI 5105: N/A
NI 5114: N/A
NI 5122: N/A
NI 5124: N/A
NI 5142: 0
NI 5152: N/A
NI 5922: N/A
Driver Prefix niScope
Driver Vendor National Instruments
Enable DC Restore 0
Enable DDC N/A
Enable Dither N/A
Enable Records Greater Than Memory0
Enable TDC0
Enable TIS NI 5105: N/A
NI 5114: N/A
NI 5122: N/A
NI 5124: N/A
NI 5142: N/A
NI 5152: FALSE
NI 5922: N/A
Enforce Realtime1
Equalization Filter EnabledNI 5105: N/A
NI 5114: N/A
NI 5122: N/A
NI 5124: N/A
NI 5142: FALSE
NI 5152: N/A
NI 5922: N/A
Equalization Num CoefficientsNI 5105: N/A
NI 5114: N/A
NI 5122: N/A
NI 5124: N/A
NI 5142: No Default
NI 5152: N/A
NI 5922: N/A
EventAny Line
Fetch Interleaved Data 0
Fetch Interleaved IQ Data NI 5105: N/A
NI 5114: N/A
NI 5122: N/A
NI 5124: N/A
NI 5142: TRUE
NI 5152: N/A
NI 5922: N/A
Fetch Meas Num Samples –1
Fetch Num Records –1
Fetch Offset 0
Fetch Record Number 0
Fetch Relative To Pretrigger
Filter Center Frequency 1000000
Filter Cutoff Frequency 1000000
Filter IIR Order 2
Filter Ripple 0.1
Filter Taps 25
Filter Type Lowpass
Filter Width 1000
FIR Filter Window 0
FIR Filter Window 0
Frequency Translation Enabled NI 5105: N/A
NI 5114: N/A
NI 5122: N/A
NI 5124: N/A
NI 5142: TRUE
NI 5152: N/A
NI 5922: N/A
Frequency Translation Phase I NI 5105: N/A
NI 5114: N/A
NI 5122: N/A
NI 5124: N/A
NI 5142: 0
NI 5152: N/A
NI 5922: N/A
Frequency Translation Phase Q NI 5105: N/A
NI 5114: N/A
NI 5122: N/A
NI 5124: N/A
NI 5142: 90
NI 5152: N/A
NI 5922: N/A
Hysteresis Percent 2
Input Clock Source No Source
Input Impedance 1000000
Interchange Check 0
Interp. Sampling Factor 2
Last ACQ Histogram Size 256
Line Number 0
Logical Name
Manufacturer National Instruments
Master Enable 0
Max Input Frequency 0, the driver selects the default bandwidth
NI 5105: 24 MHz
NI 5114: 125 MHz
NI 5122: 35 MHz
NI 5124: 60 MHz
NI 5142: 35 MHz
NI 5152: 300 MHz
NI 5922: varies based on sample rate
Max Realtime Sampling Rate NI 5105: 60 MS/s
NI 5114: 250 MS/s
NI 5122: 100 MS/s
NI 5124: 200 MS/s
NI 5142: 100 MS/s
NI 5152: 1 GS/s
NI 5922: 15 MS/s
Max RIS Rate NI 5105: N/A
NI 5114: 5 GS/S
NI 5122: 2 GS/s
NI 5124: 4 GS/s
NI 5142: 2 GS/s
NI 5152: 20 GS/s
NI 5922: N/A
Min Number of Points 1000
Min Sample Rate 1000000
Model
Other Channel 0
Output Clock Source No Source
OSP Overflow Error Reporting NI 5105: N/A
NI 5114: N/A
NI 5122: N/A
NI 5124: N/A
NI 5142: Warning
NI 5152: N/A
NI 5922: N/A
Percentage Method Base Top
Percent Waveform Transient 20
PLL Locked 0
Points Done 0
Polarity Negative
Poly. Interp. Order 1
Probe Attenuation 1
Prog FIR Filter Taps N/A
Q Source NI 5105: N/A
NI 5114: N/A
NI 5122: N/A
NI 5124: N/A
NI 5142: The channel that the attribute is registered to
NI 5152: N/A
NI 5922: N/A
Record Arm Source No Source
Records Done 0
Ref Trigger Detector Location NI 5105: N/A
NI 5114: N/A
NI 5122: N/A
NI 5124: N/A
NI 5142: Analog Detection Circuit
NI 5152: N/A
NI 5922: N/A
Ref Trig Min Quiet Time NI 5105: N/A
NI 5114: N/A
NI 5122: N/A
NI 5124: N/A
NI 5142: 0
NI 5152: N/A
NI 5922: N/A
Reference Clock Rate 10 MHz
Reference Level Units Percentage
Reference Position 50
Resolution NI 5105: 12
NI 5114: 8
NI 5122: 14
NI 5124: 12
NI 5142: 14
NI 5152: 8
NI 5922: 22
Resource Descriptor device/session-specific
RIS Method Exact Num Avg
RIS Num Avg 4
Sample Clock Output Terminal None
Sample Clock Timebase Divisor 1
Sample Clock Timebase Rate NI 5105: 60 MHz
NI 5114: 250 MHz
NI 5122: 100 MHz
NI 5124: 200 MHz
NI 5142: 100 MHz
NI 5152: 1 GHz
NI 5922: 15 MHz
Sample Clock Timebase Source No Source
Sample Mode Real-Time
Signal Format NTSC
Slave Trigger Delay 0
Start Trigger (Acq. Arm) Source Immediate
Supported Instrument Models
SYNCOUT Clock Select N/A
Temperature N/A
Test Mode Sin/Cos N/A
Time Hist. High Time 0.0005
Time Hist. High Volts 10
Time Hist. Low Time–0.0005
Time Hist. Low Volts–10
Time Hist. Size 256
Time Per Record 0.001
Trigger Coupling DC
Trigger Delay0
Trigger From PFI Delay 0
Trigger From RTSI Delay 0
Trigger From Star Delay 0
Trigger Holdoff 0
Trigger Hysteresis 0.25
Trigger Impedance 1000000
Trigger Level 0
Trigger Output Event None
Trigger Output Source No Source
Trigger Slope Positive
Trigger Source No Source
Trigger To PFI Delay 0
Trigger To RTSI Delay 0
Trigger To Star Delay 0
Trigger Type Immediate
Trigger Window High Level 0
Trigger Window Low Level 0
Trigger Window Mode Entering
Vertical Coupling DC
Vertical Offset 0
Vertical Range 10.0
V. Hist. High Volts 10
V. Hist. Low Volts –10
V. Hist. Size 256