niScope Cal Adjust Offset Range
For the NI 5132/5133, this function externally calibrates the vertical offset. Refer to the NI 5132/5133 Calibration Procedure for the appropriate stimulus voltages.
Note Use this VI only when following the NI 5132/5133 Calibration Procedure. |
instrument handle is the instrument handle that you obtain from niScope Cal Start. The handle identifies a particular instrument's session. | |||||||
channel name is the channel to calibrate. For more information, refer to Channel String Syntax. | |||||||
range (V) is the vertical range to calibrate. | |||||||
stimulus is the voltage of the applied DC signal. | |||||||
error in describes error conditions that occur before this VI runs. The default input of this cluster is no error. If an error already occurred, this VI returns the value of error in in error out. The VI runs normally only if no incoming error exists. Otherwise, the VI passes the error in value to error out. The error in cluster contains the following parameters:
| |||||||
instrument handle out has the same value as the instrument handle. | |||||||
error out contains error information. If error in indicates that an error occurred before this VI or function ran, error out contains the same error information. Otherwise, it describes the error status that this VI or function produces.
|