Waveform Measurement:Voltage Histogram:Low Volts

NI-SCOPE LabVIEW

Waveform Measurement:Voltage Histogram:Low Volts

Short Name: V. Hist. Low Volts

The last-acquisition histogram method is used, where the voltage low result is the voltage of the histogram bin with the maximum number of hits below 40% of the waveform's voltage peak-to-peak value. This calculation is useful for ignoring the overshoot and preshoot on square waves.

Units

Volts

Related Topics

Voltage Extrema Measurements