Extends the maximum sample rate on the specified Active Channel for devices that support Time Interleaved Sampling (TIS). TIS enables the device to use multiple ADCs to sample the same waveform at a higher effective real-time rate.

Default Value: FALSE (0)

Defined Values

TRUE (1)—Use multiple interleaved ADCs to acquire data for this channel.

FALSE (0)—Use only this channel's ADC to acquire data for this channel.

Remarks

The following table lists the characteristics of this property.

Data Type ViBoolean
Access R/W
Channel Based Yes