Triggering:Trigger Window:Trigger Window Low Level

NI-SCOPE LabVIEW

Triggering:Trigger Window:Trigger Window Low Level

Short Name: Trigger Window Low Level

Pass the lower voltage threshold you want the digitizer to use for window triggering.

The digitizer triggers when the trigger signal enters or leaves the window you specify with this property and Trigger Window High Level.

Valid Range: The values of the Vertical Range and Vertical Offset properties determine the valid range for this property on the channel you specify with the Trigger Source.

The value you pass for this parameter must meet the following conditions:

Low Trigger Level <= Vertical Range/2 + Vertical Offset

Low Trigger Level >= (–Vertical Range/2) + Vertical Offset

Low Trigger Level < High Trigger Level

Remarks

The following table lists the characteristics of this property.

Data Type ViReal64
Access R/W
Channel Based No