niFgen Cal Adjust Oscillator Frequency
Calculates calibration constants pertaining to the oscillator. During external calibration, you can generate sine waves and take measurements of the resulting output frequencies. The desired and measured frequencies are passed to this VI so that NI-FGEN can calculate the appropriate calibration constants and, when the calibration session is committed, store them in the onboard EEPROM.
Instrument Handle identifies your instrument session. Instrument Handle was obtained from the niFgen Init Ext Cal VI. | |||||||
Desired Frequency In Hz specifies the expected frequency of the output waveform. | |||||||
Measured Frequency in Hz specifies the actual (measured) frequency of the output waveform. | |||||||
error in describes error conditions that occur before this VI or function runs.
| |||||||
Instrument Handle Out passes a reference to your instrument session to the next VI. | |||||||
error out contains error information. If error in indicates that an error occurred before this VI or function ran, error out contains the same error information. Otherwise, it describes the error status that this VI or function produces.
|