niFgen Cal Adjust Main Path Output Impedance
Calculates calibration constants pertaining to main analog path output impedance. During external calibration, you can put the device in different configurations and take measurements of the resulting output voltage across different loads. The configuration data, as well as the measurements, are passed to this VI so that NI-FGEN can calculate the appropriate calibration constants and, when the calibration session is committed, store them in the onboard EEPROM.
Instrument Handle identifies your instrument session. Instrument Handle was obtained from the niFgen Init Ext Cal VI. | |||||||
Channel Name specifies the channel this VI uses. | |||||||
Measured Source Voltage specifies the analog output voltage measured across a very high-impedance load. | |||||||
Load Impedance specifies the impedance of the load across which the measurement passed in as Measured Voltage Across Load is taken. | |||||||
Configuration specifies the configuration of the device for this stage of calibration. | |||||||
Measured Voltage Across Load specifies the analog output voltage measured across the load impedance specified in Load Impedance. | |||||||
error in describes error conditions that occur before this VI or function runs.
| |||||||
Instrument Handle Out passes a reference to your instrument session to the next VI. | |||||||
error out contains error information. If error in indicates that an error occurred before this VI or function ran, error out contains the same error information. Otherwise, it describes the error status that this VI or function produces.
|