DAQmx Calibration Info Properties

NI-DAQmx Properties

DAQmx Calibration Info Properties

The DAQmx Calibration Info properties return information about the calibration of a device.

PropertyDescription
Active Device Specifies the device from which to retrieve calibration properties. Details
Self Calibration:Is Supported Indicates whether the device supports self calibration. Details
Self Calibration:Last Self Calibration Date/Time Indicates the last date and time that the device underwent a self calibration. Details
Self Calibration:Last Self Calibration Temperature Indicates in degrees Celsius the temperature of the device at the time of the last self calibration. Compare this temperature to the current onboard temperature to determine if you should perform another calibration. Details
External Calibration:Recommended Interval Indicates in months the National Instruments recommended interval between each external calibration of the device. Details
External Calibration:Last External Calibration Date/Time Indicates the last date and time that the device underwent an external calibration. Details
External Calibration:Last External Calibration Temperature Indicates in degrees Celsius the temperature of the device at the time of the last external calibration. Compare this temperature to the current onboard temperature to determine if you should perform another calibration. Details
User-Defined Information:Information Specifies a string that contains arbitrary, user-defined information. This number of characters in this string can be no more than Cal.UserDefinedInfo.MaxSizeDetails
User-Defined Information:Max Size Indicates the maximum length in characters of Cal.UserDefinedInfoDetails
More:Device Temperature Indicates in degrees Celsius the current temperature of the device. Details