Digitizer Sample Clock Timebase Rate

NI RF Vector Signal Analyzers

Clocking:Digitizer Sample Clock Timebase Rate

Short Name: DigitizerSampClkTimebaseRate

Specifies the frequency, in hertz, of the external clock used as the timebase source if the Digitizer Sample Clock Timebase Source is an external source.

If timebase rate is set to a value below 60 MHz, signals at frequencies just above the 20 MHz passband of the downconverter may be aliased back into the passband. This aliasing occurs because the IF frequency of the downconverter is at 15 MHz, and the upper end of the passband is at 25 MHz. At sampling rates below 60 MHz, the Nyquist frequency is close to the end of the passband and creates aliases that are not effectively filtered by the downconverter.

Remarks

The following table lists the characteristics of this property.


Data Type ViReal64
Permissions R/W
High-Level VI None