NI 5672 IQ Rates

NI RF Signal Generator

NI 5672 IQ Rates

In Arb waveform and Script modes, the NI 5672 supports several IQ rates:

  • 100 megasamples per second (MS/s) (default)
  • 50 MS/s
  • (100 MS/s)/n, where n can be of the values expressed in the following table:

    nIn steps of
    1N/A
    2N/A
    4-102
    12-4,0964
    4,096-8,1928
    8,192-16,38416

Refer to the IQ Rate (S/s) property or the NIRFSG_ATTR_IQ_RATE attribute for more information about IQ rates.

Note Note  100 and 50 MS/s IQ rates are only supported for backwards compatibility with applications written to use the NI 5670/5671. For all other applications, use an IQ rate less than or equal to 25 MS/s.