NI 5672 IQ Rates
In Arb waveform and Script modes, the NI 5672 supports several IQ rates:
- 100 megasamples per second (MS/s) (default)
- 50 MS/s
- (100 MS/s)/n, where n can be of the values expressed in the following table:
n In steps of 1 N/A 2 N/A 4-10 2 12-4,096 4 4,096-8,192 8 8,192-16,384 16
Refer to the IQ Rate (S/s) property or the NIRFSG_ATTR_IQ_RATE attribute for more information about IQ rates.
Note 100 and 50 MS/s IQ rates are only supported for backwards compatibility with applications written to use the NI 5670/5671. For all other applications, use an IQ rate less than or equal to 25 MS/s. |