Output Enabled

NI RF Signal Generator

Output Enabled

You can disable signal output at the upconverter module RF OUTPUT front panel connector using the Output Enabled property or the NIRFSG_ATTR_OUTPUT_ENABLED attribute. You can change the output enable state at any time during waveform generation, and the generation continues on internally.

Refer to the Output Enabled property or the NIRFSG_ATTR_OUTPUT_ENABLED attribute for more information about enabling signal output.

 Note   The output enabling circuitry uses mechanical relays and attenuators to switch between the enabled and disabled states. When you change a setting that results in asserting new relay or attenuator settings, the output signal may be interrupted for up to 10 ms as electromechanical components debounce.