Output Enabled

From NI RF Signal Generator

Output Enabled

You can disable signal output at the upconverter module RF OUTPUT front panel connector using the Output Enabled property or the NIRFSG_ATTR_OUTPUT_ENABLED attribute. You can change the output enable state at any time during waveform generation, and the generation continues on internally.

Refer to the Output Enabled property or the NIRFSG_ATTR_OUTPUT_ENABLED attribute for more information about enabling signal output.

 Note   The output enabling circuitry uses mechanical relays and attenuators to switch between the enabled and disabled states. When you change a setting that results in asserting new relay or attenuator settings, the output signal may be interrupted for up to 10 ms as electromechanical components debounce.